7.2
High Voltage Output Calibration
Use S1 and S2 to adjust the voltage, and please adjust the value of CODE and
READING the same if possible. After the adjustment, press ENTER to move next time
UP:
Press
Figure 7.2.1
to increase CODE value
DOWN:
Press
SAVE:
Press
to decrease CODE value
to save CODE
DEFAULT:
Press
to reset default
QUIT:
Press
to quit the calibration
7.3
High Voltage Output Testing
Connect the positive terminal of DC meter to CH7 of the instrument and negative
terminal to CH8. The result should be minus value due to the voltage is negative while
testing
7.3.1
High Voltage Output Setting
Press FUNC under the main menu to enter System mode, and then press S3 to
enter H/V Test mode
S1
S4