MICROTEST
20
Figure 5.5.2.2
Voltage:
Move the cursor to Voltage and use numeric key to set up voltage value, the
maximum voltage is 1.0kV, the minimum voltage is 0.1kV
Dwell:
High voltage output sequential times. Use numeric key to set up, maximum
value is 99.9 second, minimum value is 0.1 second
Max/Min:
Use numeric key to set up values, the maximum value is 9999MΩ,
the minimum value is 0MΩ
Arc Sen.:
The smaller value, the higher sensitivity. Use numeric key to set up
values, if value is 0 then the function is off.
DUT No.:
Use numeric key to set up, the maximum value is 4, the minimum value
is 0. Use 0 at the first step, after the first step the minimum value should
be 1. This function is to make users speed up testing.