Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment - guidance
Voltage dips
(IEC
61000-4-11)
0%
U
T
For 0.5 cy-
cle
A 0°, 45°,
90°, 135°,
180°, 225°,
270° and
315°
0%
U
T
for 1
cycle
and
70%
U
T
for
25 cycles,
50Hz
30 cycles,
60 Hz,
Monophase:
0°
Non applica-
ble
Mains power quality should be that of a
Home health care environment or profes-
sional health care establishment.
If operation of the system requires continued
use during power cuts, it is recommended
that the medical device be powered by a
separate power source (UPS, etc.).
Voltage Inter-
ruptions
(IEC
61000-4-11)
0 %
U
T
;
For 250 cy-
cles, 50 Hz
For 300 cy-
cles, 60 Hz
Non applica-
ble
Mains power quality should be that of a
Home health care environment or profes-
sional health care establishment
If operation of the system requires continued
use during power cuts, it is recommended
that the medical device be powered by a
separate power source (UPS, etc.).
NOTE:
U
T
is the a.c. mains voltage prior to application of the test level.
WARNING:
Portable RF communication devices (including peripherals such as antenna
cables and external antennae) should not be used closer than 30 cm (12 inches) to any part
of the SpiderView, including the specified cables. Otherwise, the performance of this device
may be impaired.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic envi-
ronment - guidance
Electromagnetic fields
Radiated RF
(IEC 61000-4-3)
10 V/m
80 MHz –
2,7 GHz
80 % MA
at 1 kHz
10 V/m
80 MHz –
2,7 GHz
80 % MA at
1 kHz
Home health care environ-
ment and professional
health care establishment.
15. APPENDIX