button to select the “BarCode” symbol to 

input with the built-in barcode scanner. 

Please refer to the user manual for more 

information.                                                                          

                                                     

 5.  Confirm the code and lot number of the 

control.

    Press “OK” button to confirm the code 

and lot number of the control. Once the 

lot number is entered, the meter will 

automatically check whether the control 

is expired. If the control is expired, the 

user shall use another control within the 

expiration period to perform the test.           

                                                     

    

Note:

 Always match the control code, lot 

number on the display with those on the 

control vials. Failure to do so may cause 

inaccurate results.

 

6.  Enter the code and lot number of the strip.

    
    
    

  

    Use the same method as indicated in step 

4 to enter the code and lot number of the 

strip printed on the label of the foil pouch. 

7. Confirm the code and lot number of the 

strip.

    Press “OK” button to confirm the code 

and lot number of the strip. Once the lot 

number of strip is entered, the meter will 

automatically check whether the strip is 

expired or not. If the strip is expired, the 

user shall use another strip within the 

expiration period to perform the test. 

                                                               

                                               

Note: 

Always match the strip code, lot 

number on the display with those on the 

label of strip. Failure to do so may cause 

unacceptable results.

8. After confirming the code and lot number 

of the strip, the meter will warm up 

automatically for the test. When it is ready 

to perform a test, the meter will beep and 

prompt the user to apply control sample.

9.   Perform the control test. 

     Before using, slightly swirl the target 

control at least 10 times and then apply 

one drop of the liquid control to the 

sample well, the system will start test 

automatically. 

Summary of Contents for qLabs QSV-1-CL Pro

Page 1: ...rocedure may alter performance 3 Take all necessary precautions required when handling laboratory reagents 4 Do not freeze liquid controls 5 Do not use expired liquid controls 6 Do not use for any purpose other than described in the Intended Use section 7 Do not interchange caps as contamination may occur and compromise results qLabs Vet Coag Panel 2 Liquid Control For Health Care Professional Use...

Page 2: ... qLabs Vet Coag Panel 2 Test Strip qLabs Vet Coag Panel 2 Liquid Control Level 1 and Level 2 Note Please refer to the user s manual for appropriate operation instructions about the test strip and the meter 2 Insert the strip Press the power button for 2 seconds to turn on the qLabs Vet Remove a new qLabs Vet Coag Panel 2 Test Strip from foil pouch Insert the strip electrode end first into the test...

Page 3: ...the strip printed on the label of the foil pouch 7 Confirm the code and lot number of the strip Press OK button to confirm the code and lot number of the strip Once the lot number of strip is entered the meter will automatically check whether the strip is expired or not If the strip is expired the user shall use another strip within the expiration period to perform the test Note Always match the s...

Page 4: ...tem is working properly and all testing procedure has been done correctly when the test results obtained fall within the assigned control PT and APTT range on the vial Note Please refer to the assigned PT and APTT control range printed on the label of the vial UNEXPECTED RESULTS If the test results fall outside of the assigned range or qLabs ElectroMeter displays error code E XXX check for the fol...

Page 5: ... 0 48 2 76 Level 2 31 97 1 12 3 49 Within Run N 20 APTT Value Mean S D CV Level 1 113 03 4 35 3 85 Level 2 144 41 4 53 3 14 Between Run N 40 APTT Value Mean S D CV Level 1 115 48 3 86 3 34 Level 2 143 80 7 03 4 89 ADDITIONAL INFORMATION For additional information about this product please call your local distributor or call our customer service at 86 755 86296766 ...

Page 6: ...Symbols Explanation In Vitro Diagnostics Name and Address of Manufacturer Temperature Limitation Lot Number Expiry Date Catalogue Number Catains Sufficient for n Tests Caution Read Carefully ...

Page 7: ...6 755 86673903 Obelis SA Bd General Wahis 53 1030 Brussels Belgium www obelis net Tel 32 2 732 59 54 Fax 32 2 732 60 03 2015 Micropoint Biotechnologies Inc All rights reserved Specifications are subject to change without notice Printed in China qLabs and Micropoint are registered trademarks of Micropoint Bioscience Inc P N 99030545431 Rev A1 ...