5.1 General use
5.2 Program Mode
① Turn on the Power switch of implant engine controller.
② Programs are selected in turn by pressing Foot switch or Program button of implant engine controller.
③ Check the displayed Torque, Speed, Coolant, Gear and Direction of rotation.
④ It starts rotating upon pressing Foot switch. It rotates at low speed with light press on Foot switch
and rotates at full speed with hard press. When water injection status is preset,
Pump starts rotating as well.
( For the convenience of the operator and the safety of the patient, a slow start function is installed
that gradually reaches the set rotation value in 2-3 seconds)
⑤ When load reaches the maximum value of preset torque during use, motor stops rotating.
⑥ It stops rotating upon releasing Foot switch.
A user selects a program necessary for surgery with this button.
Program alters in turn from number 1 to 6 upon pressing Program
button each time. It alters in order of Drilling→Tapping→Remove Tap
→Implant→Remove→Lock screw.
Cursor locates on border of the selected program.
Program
Button
Chapter 5. Operation
5.2.1 Selecting a program
This function deletes an unnecessary and unused programs.
Selects an unused program and Deletes pressing this button more
than 3 seconds.
Program
Button
5.2.2 Deleting a program
This function restores the deleted programs.
Restores all deleted programs pressing this button more than 2 seconds.
Memory
Button
5.2.3 Restoring programs