Page 52
UF AV5500 Area-Velocity Flow Meter
Register
Address
Description
Register
Type
Format Type
Comments
30327
Previous day Volume
Total - IG
Input Register
Floating Point Register (1 of 2)
IG = Imperial Gallons
30328
Previous day Volume
Total - IG
Input Register
Floating Point Register (2 of 2)
IG = Imperial Gallons
30329
Previous day Volume
Total - IMG
Input Register
Floating Point Register (1 of 2)
IMG = Imperial Million
Gallons
30330
Previous day Volume
Total - IMG
Input Register
Floating Point Register (2 of 2)
IMG = Imperial Million
Gallons
30331
Previous day Volume
Total - bbl
Input Register
Floating Point Register (1 of 2)
bbl = US Oil Barrel = 42
Gallons
30332
Previous day Volume
Total - bbl
Input Register
Floating Point Register (2 of 2)
bbl = US Oil Barrel = 42
Gallons
30501
Sensor Range - inches
Input Register
Floating Point Register (1 of 2)
Only used when through-air
sensor is installed.
30502
Sensor Range - inches
Input Register
Floating Point Register (2 of 2)
Only used when through-air
sensor is installed.
30503
Sensor Range - feet
Input Register
Floating Point Register (1 of 2)
Only used when through-air
sensor is installed.
30504
Sensor Range - feet
Input Register
Floating Point Register (2 of 2)
Only used when through-air
sensor is installed.
30505
Sensor Range - mm
Input Register
Floating Point Register (1 of 2)
Only used when through-air
sensor is installed.
30506
Sensor Range - mm
Input Register
Floating Point Register (2 of 2)
Only used when through-air
sensor is installed.
30507
Sensor Range - meters
Input Register
Floating Point Register (1 of 2)
Only used when through-air
sensor is installed.
30508
Sensor Range - meters
Input Register
Floating Point Register (2 of 2)
Only used when through-air
sensor is installed.
30525
Sensor Level - inches
Input Register
Floating Point Register (1 of 2)
30526
Sensor Level - inches
Input Register
Floating Point Register (2 of 2)
30527
Sensor Level - feet
Input Register
Floating Point Register (1 of 2)
30528
Sensor Level - feet
Input Register
Floating Point Register (2 of 2)
30529
Sensor Level - mm
Input Register
Floating Point Register (1 of 2)
30530
Sensor Level - mm
Input Register
Floating Point Register (2 of 2)
30531
Sensor Level - meters
Input Register
Floating Point Register (1 of 2)
30532
Sensor Level - meters
Input Register
Floating Point Register (2 of 2)
30709
Flow Temperature - F
Input Register
Floating Point Register (1 of 2)
30710
Flow Temperature - F
Input Register
Floating Point Register (2 of 2)
Summary of Contents for UF AV5500
Page 2: ...Note This page has been left blank intentionally...
Page 28: ...Page 28 UF AV5500 Area Velocity Flow Meter...
Page 34: ...Page 34 UF AV5500 Area Velocity Flow Meter Resistance vs Temperature...
Page 40: ...Page 40 UF AV5500 Area Velocity Flow Meter SS PIPE MOUNTING BAND OPTION VSJ...
Page 61: ...Page 61 UF AV5500 Area Velocity Flow Meter Dual Non Contacting Sensor Configuration...