
Page 23
UF AV5500 Area-Velocity Flow Meter
the calibration menus.
Press
✓
to return to
Menu Selections
.
SIMULATION
Exercises the 4-20mA (0-5V) outputs, digital display and control relays.
Test
Select
Maximum
and press
✓
to simulate maximum Flow, Level and
Velocity and to output 20mA (5V) to the analog channels.
Select
Minimum
and press
✓
to simulate minimum Flow, Level and
Velocity and to output 4mA (0V) to the analog channels.
To simulate an intermediate Flow, Level and Velocity set
Test
to
Actual
and then enter a value for the
Level
and
Velocity
. The Flow
calculation, analog outputs and control relays will respond to the
simulated values.
Summary of Contents for UF AV5500
Page 2: ...Note This page has been left blank intentionally...
Page 28: ...Page 28 UF AV5500 Area Velocity Flow Meter...
Page 34: ...Page 34 UF AV5500 Area Velocity Flow Meter Resistance vs Temperature...
Page 40: ...Page 40 UF AV5500 Area Velocity Flow Meter SS PIPE MOUNTING BAND OPTION VSJ...
Page 61: ...Page 61 UF AV5500 Area Velocity Flow Meter Dual Non Contacting Sensor Configuration...