5: Troubleshooting
32
Portaflow 220 User Manual
(Issue 1.0)
4-20mA ERRORS
mA out > Max
Interpretation:
The actual flow is higher than the maximum set on the mA range.
Response:
Re-scale the 4-20mA output to be able to cope with the higher flow –
see
Paragraph 3.7.2
.
Calibration 20mA
Error!
NOTE: The 4-20mA output is calibrated before the instrument leaves the factory and
should not require further adjustment.
Interpretation:
You have adjusted the DAC outside its accepted range when
calibrating the 20mA signal output.
Response:
Re-calibrate the 4-20mA output – see
Paragraph 3.7.2
.
Calibration 4mA Error!
NOTE: The 4-20mA output is calibrated before the instrument leaves the factory and
should not require further adjustment.
Interpretation:
You have adjusted the DAC outside its accepted range when
calibrating the 4mA signal output.
Response:
Re-calibrate the 4-20mA output – see
Paragraph 3.7.2
.
SET-UP ERRORS
Pipe OD out of range
Interpretation:
You have entered an out-of-range value for the pipe outside
diameter dimension – i.e. larger or smaller than the unit or sensor can be used on.
Response:
Enter a valid number.
Wall thickness out of
range
Interpretation:
You have entered an out-of-range value for the pipe wall thickness
dimension – accepted range is 1mm - 75mm.
Response:
Enter a valid number.
Lining thickness out of
range
Interpretation:
You have entered an out-of-range value for the lining thickness
dimension – acceptable range is 0mm - 10mm.
Response:
Enter a valid number.
Temperature range
Interpretation:
You have entered an out-of-range value for the fluid Temperature.
Accepted temperature range -20°C to +300°C.
Response:
Enter a valid number.
Invalid Date or Time
Interpretation:
The entered Date or Time is invalid, or when setting up ‘timed’ data
logging the Stop time is set earlier than the Start time.
Response:
Enter a valid Date and Time.
Sensors: INVALID
Interpretation:
The selected temperature is higher than the maximum allowed for
the sensor type.
Response:
Enter a different temperature.
Mode: Err Typ
Interpretation:
The selected sensors are invalid and the mode cannot be verified.
Response:
Select a mode that gives a non-zero separation distance.
Summary of Contents for portaflow 220
Page 2: ......
Page 14: ...2 Installation 10 Portaflow 220 User Manual Issue 1 0...
Page 30: ...3 Operating Procedures 26 Portaflow 220 User Manual Issue 1 0...
Page 32: ...4 Maintenance Repair 28 Portaflow 220 User Manual Issue 1 0...
Page 42: ...A Specification 38 Portaflow 220 User Manual Issue 1 0...
Page 43: ...A Specification Portaflow 220 User Manual 39 Issue 1 0 CE Declaration of Conformity...
Page 44: ...A Specification 40 Portaflow 220 User Manual Issue 1 0...