Micronics portaflow 220 User Manual Download Page 26

3:  Operating Procedures

22

Portaflow 220 User Manual

  (Issue 1.0)

3.6

Performing Monitoring Functions

3.6.1  How to measure totalised flows (manually)

The basic measurement indicated on the 

FLOW READING

 screen is the instantaneous flow rate, which in 

some applications may vary over a period of time. Average flow rates are therefore often required in order to 
get a better understanding of an application’s true performance. This is simply achieved by noting the total 
flow over a specific period (for example 30-60 minutes) and then calculating the average flow rate over that 

period of time.

Calculating the average flow

To calculate the average flow wait for the allotted monitoring period to expire then divide the indicated total 
flow by the time taken. This will give you the average flow in m/s, galls/hours or whatever units you select.

Note that in a bi-directional flow situation you must take the difference between the indicated positive and 
negative flow totals before carrying out the average flow rate calculation.

How to stop the totaliser temporarily

If, for operational reasons, you want to stop the totaliser function temporarily set the 

Totaliser

 option to 

Stall

 in the 

FLOW READING OPTIONS

 screen as described above. This will stop the totaliser operation 

without affecting its current values.

1.

Press the 

Options

 key to access the 

FLOW 

READING OPTION

 screen shown.

2.

If the 

Totaliser

 is indicating 

Run

, select it 

and change it to 

Stall

. Press 

ENTER.

3.

Select 

Reset +Total

 and press 

ENTER

.

4.

Press 

ENTER

 on the subsequent screen to 

accept the reset.

5.

Press 

ENTER

 again to return to the 

FLOW 

READING

 

OPTIONS

 menu.

FLOW READING OPTION 

 

DD-MM-YY  HH:MM:SS

                          

    

 Data review

 Zero Cutoff (m/s)     :        0.00

 Set zero flow (m/s)   :        0.00

 Damping (secs)        :        10

 Totaliser             :       Stall

 Reset +Total

6.

Select 

Reset –Total

 and press 

ENTER

.

7.

Press 

ENTER

 on the subsequent screen to 

accept the reset.

8.

Press 

ENTER

 again to return to the 

FLOW 

READING

 

OPTIONS

 menu.

9.

Note and record the current time.

10. Select 

Totaliser

 and change it to 

Run

 

then press 

ENTER.

Note: the totalisers begin to count up as 
soon as

 Totaliser

 is put to 

Run

.

FLOW READING OPTION 

 

DD-MM-YY  HH:MM:SS

                          

    

Zero Cutoff (m/s)      :        0.00

 Set zero flow (m/s)   :        0.00

 Damping (secs)        :        10

 Totaliser             :       Run

 Reset +Total

 Reset –Total

11. Scroll down and select 

Exit

 to return to the 

FLOW READING

 screen which will now 

indicate the instantaneous flow together 
with the totalised flow.
Note that in some installation the measured 
flow can be in either direction. Where this is 
the case the upstream flow is shown 
separately in the 

–Total

 field.

FLOW READING 

 

DD-MM-YY  HH:MM:SS

                          

    Dim: mm

Signal

                               l/min

      +Total: 300.0         litres

      –Total: 0.00          litres

12.34

Summary of Contents for portaflow 220

Page 1: ...ic Flowmeter User Manual Micronics Ltd Knaves Beech Business Centre Davies Way Loudwater High Wycombe Bucks HP10 9QR Telephone 44 0 1628 810456 Facsimilie 44 0 1628 531540 E mail sales micronicsltd co...

Page 2: ......

Page 3: ...3 3 2 Using the Quick Start Menu 13 3 3 Using the System at a Regularly Monitored Location 16 3 4 Managing Named Sites 17 3 4 1 Setting up a new site 18 3 4 2 Changing a site name 18 3 5 Instrument Ca...

Page 4: ...Maintenance Repair 27 5 Troubleshooting 29 5 1 Overview 29 5 2 General Troubleshooting Procedure 30 5 3 Warning Status Messages 31 5 4 Test Block 33 5 5 Microprocessor Reset Facility 34 5 6 Diagnosti...

Page 5: ...of pipe diameters The PF220A can be used with pipes in the range 13mm 115mm and the PF220B with pipes in the range 50mm 1000mm Easy to operate the Portaflow 220 standard features are Large easy to rea...

Page 6: ...ing the correct separation distance between the transducers to obtain the strongest signal Figure 1 1 Operating modes Fluid flow Fluid flow Fluid flow Fluid flow Fluid flow U U U U D D D D Separation...

Page 7: ...ith UK US European adaptors 110 240VAC 4 20mA Pulse Output cable USB cable and RS232 C cable 2 lengths of chain Test block Transducer cables x2 2 metres long one red and one blue Transducer set x2 typ...

Page 8: ...be connected to the RED socket and the downstream transducer to the BLUE one for a positive flow reading It is safe to connect or disconnect the cable while the instrument is switched on 4 20mA and Pu...

Page 9: ...indicate the active menu choice which can then be selected by pressing the ENTER SELECT key Some menus have more options than can be shown on the screen at the same time in which case the overflowed...

Page 10: ...eading mode the percentage battery charge level is displayed symbolically on the LCD screen A warning message is triggered if the charge falls to approximately 30 at which point there is up to four ho...

Page 11: ...g errors can be unpredictable Key Point Do not expect to obtain accurate results if the transducers are positioned close to any obstructions that distort the uniformity of the flow profile Flow Valid...

Page 12: ...fected 2 Prepare the pipe by degreasing it and removing any loose material or flaking paint in order to obtain the best possible surface A smooth contact between pipe surface and the face of the trans...

Page 13: ...ver plate A fully towards the outside of the guide assembly to allow sufficient access to fit the transducer 2 Clean the face of the transducer removing all traces of dirt and grease 3 Apply a 3mm bea...

Page 14: ...2 Installation 10 Portaflow 220 User Manual Issue 1 0...

Page 15: ...o Flow Offset Paragraph 3 5 2 How to adjust the Calibration Factor Paragraph 3 5 3 How to adjust the Roughness Factor Paragraph 3 5 4 How to adjust the Damping Factor Paragraph 3 5 5 Set up a monitori...

Page 16: ...lected language will be the default when the instrument is next used To change the language again select the Change Language option in the SETUP INSTRUMENT screen see below 4 The initialisation screen...

Page 17: ...a Site within the Portaflow 220 which then stores the site parameters See Paragraph 3 4 Before you can use the Portaflow system you need to obtain the following details this information will be requir...

Page 18: ...the list provided then press ENTER If the material is not listed select Other and enter the propagation rate of the pipe wall material in metres sec Contact Micronics if this is not known PIPE WALL MA...

Page 19: ...ct either C or F with the cursor and press the ENTER key 12 Enter the new temperature value and press the ENTER key 13 The new temperature should now be indicated in both C and F 14 Select Continue an...

Page 20: ...e a problem with the transducer spacing alignment or connections or it could be due to an application problem FLOW READING DD MM YY HH MM SS Dim mm Signal l min Total 0 00 litres Total 0 00 litres 1 S...

Page 21: ...the Portaflow 220 system These can then be recalled later when revisiting that particular location The instrument can store up to 20 sites the first site is reserved for QUICK START and cannot be rena...

Page 22: ...message 7 On completion press ENTER then select Exit This will take you back to the VIEW EDIT SITE DATA screen ALTER NAME DD MM YY HH MM SS Site name EmptySite1 Do you wish to Alter the Site Name Exi...

Page 23: ...w 2 With the instrument in FLOW READING mode press the Velocity function key and observe the reading m s Any reading other than 0 000 indicates an offset error and in practice this will typically be i...

Page 24: ...tracted from the flow reading across the whole range Key Point USE THIS FACILITY WITH CARE ONLY WHERE NECESSARY The Portaflow instrument is fully calibrated before leaving the factory and under nor ma...

Page 25: ...factor according to the pipe material and condition as described above 4 Press Enter to apply the change FLOW READING OPTION DD MM YY HH MM SS Data review Zero Cutoff m s 0 00 Set zero flow m s 0 00...

Page 26: ...described above This will stop the totaliser operation without affecting its current values 1 Press the Options key to access the FLOW READING OPTION screen shown 2 If the Totaliser is indicating Run...

Page 27: ...Output or select one of the output ranges to turn it ON 5 Press ENTER to return to the 4 20mA OUTPUT screen 4 20 mA OUTPUT DD MM YY HH MM SS Dim mm Off 4 20mA 0 20mA 0 16mA Key Point The 4 20mA outpu...

Page 28: ...the Scroll keys to obtain an output of exactly 20mA The DAC should indicate approximately 40000 6 Press ENTER when done CALIBRATE 20mA DD MM YY HH MM SS Dim mm Adjust the output current to 20mA Use UV...

Page 29: ...ting in the FLOW READING mode press the Pulse function key to access the PULSE OUTPUT screen 2 A Pulse output is ON message appears in the second line of the display 3 Select Exit and press ENTER to r...

Page 30: ...3 Operating Procedures 26 Portaflow 220 User Manual Issue 1 0...

Page 31: ...d that the ultrasonic couplant is replaced on the sensors every 6 months especially on pipes where the application is too hot to touch If the signal level drops below 30 this is also an indication tha...

Page 32: ...4 Maintenance Repair 28 Portaflow 220 User Manual Issue 1 0...

Page 33: ...reventing good surface contact with the transducer Flaking paint should be removed Variable air gap in concrete covered pipes affecting the ultrasonic signal quality Poor internal pipe construction Ro...

Page 34: ...nsure correct transducer type been selected Ensure transducer separation distance correct Ensure transducers are aligned correctly Ensure pipe diameter within specifications of the selected transducer...

Page 35: ...the Options menu Response Enter a valid number Damping error Interpretation You have entered an out of range value in the Damping field in the Options menu Response Enter a valid number Totaliser bey...

Page 36: ...ue for the pipe outside diameter dimension i e larger or smaller than the unit or sensor can be used on Response Enter a valid number Wall thickness out of range Interpretation You have entered an out...

Page 37: ...rs positioned towards the centre of the test block as shown and temporarily secure them in place using elastic bands or tape 9 Connect the sensors to the Portaflow 220 instrument using the cables prov...

Page 38: ...the following parameters Calculated time s This is a value the instrument predicts will be the time in secs that it should take for the acoustic wave to propagate across a particular pipe size This va...

Page 39: ...min USgals h Barrel h Barrel day m s m min m h Selectable Volume Units l gal USgals Barrel m Total Volume 12 Digits forward and reverse APPLICABLE FLUID TYPES Fluid Condition Clean liquids or oils th...

Page 40: ...0mA ELECTRICAL Supply Voltage Input Voltage Range 9 24Vdc Power Consumption 10 5W Battery Technology 5 cell NiMH Capacity 3 8AHr Operating Time Typically 20 hours continuous with backlight and 4 20mA...

Page 41: ...erating Temperature 20 C to 50 C Storage Temperature 25 C to 65 C Operating Humidity 90 RH MAX at 50 C APPROVALS Safety BS EN 61010 EMC BS EN 61326 1 2006 BS EN 61326 2 3 2006 Battery Charger EN61204...

Page 42: ...A Specification 38 Portaflow 220 User Manual Issue 1 0...

Page 43: ...A Specification Portaflow 220 User Manual 39 Issue 1 0 CE Declaration of Conformity...

Page 44: ...A Specification 40 Portaflow 220 User Manual Issue 1 0...

Reviews: