4 D
IAGNOSTICS
Data Results on page 2 - 20
Unit Configuration on the next page
Adsorptive Line Test on page 4 - 3
Helium Line Test on page 4 - 5
System Leak Test on page 4 - 7
System Check on page 4 - 9
Zero Test on page 4 - 10
Alt + 7
Press
CHOICE
to display the options. Press
ENTER
to select.
Use to:
n
View analyzer and calibration statistics
n
Clean and verify the gas line for the adsorptive gas
n
Clean and verify the gas line for helium
n
Check the system for leaks (service request only)
n
Verify that the analyzer is operating properly (service request only)
n
Zero transducers to stored offsets (service request only)
Some of the diagnostics require operator attention and a beep sounds at that time. The beeping
continues at intervals until the operator performs the task shown in the display window. Do not
press
ENTER
until the task has been performed. After a test has completed, a report is generated
automatically to the destination(s) specified in
Report Options
.
It is not recommended that diagnostic reports be transmitted via serial line
communication. Diagnostic results in this form produce values only and will require
interpretation by a Service representative.
Choices:
Unit Configuration, Adsorptive Line Test, Helium Line Test, System Leak Test, System
Check, Zero Test
4 Diagnostics
Gemini VII Keypad Operator Manual
239-42801-01 (Rev E) — July 2021
4 - 1