Gemini VII
Preventive Maintenance
239-42828-01 - Dec 2012
9-11
Replacing the Alumina in the Oil Vapor Trap
The activated alumina in the oil vapor trap becomes saturated during use. The alumina should be
inspected periodically and replaced when most of the alumina pellets are no longer white.
1.
Disconnect the vacuum pump from the analyzer and place it on a work table.
2.
Loosen the wing nut on the clamp at the top of the oil vapor trap. Swing the clamp open and
remove the trap from the hose.
3.
Loosen the wing nut on the clamp at the bottom of the oil vapor trap. Swing the clamp open and
remove the trap.
4.
Remove one end fitting from the trap body and dispose of the used alumina in an appropriate
manner.
5.
Wash the trap body with a detergent-based soap. Rinse with water, then rinse with isopropyl or
ethyl alcohol. Set the trap aside and allow to dry thoroughly.
Do not perform the following procedure on used alumina. The resultant oil vapors may
cause a fire or an explosion.
Exposure of the trap body to oil vapor may cause small cracks on the inside surface of
the trap body. Under normal circumstances, these cracks will not cause problems or
leaks.
Summary of Contents for Gemini VII 2390a
Page 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Page 4: ......
Page 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Page 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Page 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Page 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Page 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Page 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Page 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Page 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Page 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Page 264: ......
Page 266: ......
Page 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Page 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Page 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Page 380: ...Contact Us vi Smart VacPrep Operator Manual...
Page 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...