System Commands
AccuPyc II 1340
3-26
Aug 2013
Volume
Before proceeding, place an empty cup into the sample chamber for
Cal1
(volume offset). If using an
insert, place the insert and its appropriate sample cup into the sample chamber. (Refer to
and Removing Inserts and Sample Cups
Press
ENTER
to accept
Volume
(default) and display the
next prompt.
Press
CHOICE
until the appropriate insert is displayed, then
press
ENTER
.
Choices:
None, 0.1 cm
3
(for 1 cm
3
pycnometer)
None, 3.5 cm
3
, 1.0 cm
3
(for 10 cm
3
pycnometer)
None, 35 cm
3
, 10 cm
3
(for 100 cm
3
pycnometer)
None, 650 cm
3
, 1300 cm
3
(for 2000 cm
3
pycnometer)
Yes
must be selected for the
Request Insert
prompt in
Report Options for this prompt to display.
Enter the volume of the calibration standard(s) you plan to
use in the calibration (located on the outside of the
calibration standards case).
Range:
0.1 to 999.0000 cm
3
Some pycnometers require and are shipped with multiple
standards; in this instance, the value recorded on the case is
the sum for all standards and is the one you should enter at
this prompt. Press
ENTER
.
Press
ENTER
to begin the calibration or
Alt + CLEAR
to
cancel the calibration.
The volume offset is calibrated first (
Cal1
). When the first
calibration is complete, the pycnometer beeps three times
and the next prompt is displayed.
Unit[n]
SN1234 10 cm3
X - X - X
Calibration Type?
Volume
Unit[n]
SN1234 10 cm3
Calibrate Volume
Chamber insert?
None
Unit[n]
SN1234 10 cm3
Calibrate Volume
Volume of Cal Std:
1.0000 cm3
Unit[n]
SN1234 10 cm3
Calibrate Volume
[ENTER] to start
[Escape] to cancel
Summary of Contents for AccuPyc 2 1340
Page 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Page 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Page 4: ......
Page 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Page 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Page 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Page 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Page 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Page 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Page 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Page 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Page 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Page 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Page 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Page 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...