AccuPyc II 1340
Installing the Analyzer
June 2013
2-5
3.
Purge the regulator and inlet tubing; this is important to prevent contamination of the
analysis gas.
a. Close the regulator shut-off valve; turn it fully clockwise.
b. Turn the pressure regulator control knob fully counterclockwise.
c. Slowly open the gas bottle shut-off valve; turn it counterclockwise, then quickly close
it.
d. Observe the high-pressure gauge.
If the pressure decreases, tighten the nut connecting the regulator to the gas bottle. If
the pressure is stable, proceed to step e.
e. Turn the pressure regulator control knob clockwise until the low-pressure gauge
indicates 15 psig (100 kPag).
f. Open the regulator shut-off valve; turn it counterclockwise.
g. Open the gas bottle shut-off valve; turn it counterclockwise. Flow gas for 10 to 30
seconds, then close the valve.
h. Close the regulator shut-off valve; turn it clockwise.
4.
Attach the other end of the copper tubing to the gas inlet located on the rear panel of the
pycnometer. The pycnometer is shipped with a plug in the inlet; remove and discard the
plug, or store in a safe location for future use.
Make sure the gas supply equipment is adequately vented before performing the next
step.
Gas inlet plug
Summary of Contents for AccuPyc 2 1340
Page 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Page 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Page 4: ......
Page 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Page 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Page 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Page 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Page 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Page 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Page 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Page 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Page 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Page 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Page 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Page 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...