Installing the Analyzer
AccuPyc II 1340
2-4
June 2013
Rear Panel Connections
1.
Attach an appropriate regulator to the gas supply bottle or gas line, then tighten with a
wrench. Leave the gas bottle shut-off valve closed until instructed otherwise.
2.
Attach the gas inlet line to the regulator, reducer fitting, or regulator expansion:
Multiple gases can be attached to the pycnometer using the Multigas option (refer to
Bottle Shut-Off
Valve
Shut-Off Valve
Nut
Two-Stage Pressure
Reducing Regulator
Gas Bottle
Brass Reducer
Fitting
Copper Tubing
Regulator
Shut-Off Valve
Pressure Control
Knob
It is very important to use the inlet tubing supplied with the analyzer. Gas lines made
of materials other than copper or stainless steel can cause operational problems as
well as inaccurate data.
If regulator has .....
Then .....
1/4-in outlet
Attach the reducer fitting to the outlet of the regulator shut-off/
isolation valve
Tighten the nut finger-tight, then 1-1/4 turns with a wrench.
Continue with the steps given for the 1/8-in outlet
1/8-in outlet
Insert the gas tubing into the fitting.
Make sure the tubing is seated fully inside the fitting.
Tighten the nut finger-tight.
While holding the fitting body steady, tighten the nut with a
wrench 3/4 turn.
Do not overtighten the fittings. Doing so can collapse the brass ferrule and cause a
leak.
Summary of Contents for AccuPyc 2 1340
Page 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Page 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Page 4: ......
Page 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Page 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Page 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Page 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Page 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Page 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Page 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Page 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Page 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Page 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Page 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Page 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...