Analysis Sequence
3Flex Service Manual
3-6
Aug 2013
11. Helium is evacuated from the samples and the Dewar is lowered if specified in the sample file.
12. Samples are evacuated for the amount of time specified in the sample file.
13. The Po tube is dosed and allowed to equilibrate.
14. If the pressure does not change measurably, the Psat measurement is taken. If the pressure does
change measurably, another dose may be required.