12
When the mains adapter is connected, no battery current is consumed.
9. Error Messages
If an error occurs during the measurement, the measurement is inter-
rupted and an error message, e.g. «
ERR 3
», is displayed.
* Please consult your doctor, if this or any other problem occurs
repeatedly.
1. Plug the adapter cable into the mains adapter socket
5
in the
blood pressure monitor.
2. Plug the adapter plug into the wall socket.
Error
Description Potential cause and remedy
«
ERR 1
» Signal too
weak
The pulse signals on the cuff are too weak.
Re-position the cuff and repeat the
measurement.*
«
ERR 2
» Error signal
During the measurement, error signals were
detected by the cuff, caused for instance by
movement or muscle tension. Repeat the
measurement, keeping your arm still.
«
ERR 3
» No pressure
in the cuff
An adequate pressure cannot be generated
in the cuff. A leak may have occurred. Check
that cuff and pump ball are correctly
connected and that the cuff is not fitted too
loosely. Replace the batteries if necessary.
Repeat the measurement.
«
ERR 5
» Abnormal
result
The measuring signals are inaccurate and no
result can therefore be displayed. Read
through the checklist for performing reliable
measurements and then repeat the
measurement.*
«
HI
»
Pulse or cuff
pressure too
high
The pressure in the cuff is too high (over 300
mmHg) OR the pulse is too high (over 200
beats per minute). Relax for 5 minutes and
repeat the measurement.*
«
LO
»
Pulse too low The pulse is too low (less than 40 beats per
minute). Repeat the measurement.*
)
If you think the results are unusual, please read through the
information in «Section 1.» carefully.
Error
Description Potential cause and remedy