8. Display Mode

When the device is switched on, shortly press the ON/OFF 

button 

1

 to switch to another display mode to select your 

desired display mode 

9

. There are 6 different display 

modes. The default setting is mode 1.

9. Using the Lanyard 

7

1. Thread the thinner end of the lanyard through the 

hanging hole at the rear end of the device.

2. Thread the thicker end of the lanyard through the 

threaded end before pulling it tightly.

10. Bluetooth

®

 Function

This device can be used in conjunction with a smartphone 

running the «Microlife Connected » App. The 

measurement results will be automatically transferred via 

Bluetooth

®

.

Downloading the «Microlife Connected » App

Download «Microlife Connected » App for free from 

Google Play™ (Android) or App Store (iOS) and install it on 

your smartphone.

How the Bluetooth

®

 Function works

The Bluetooth

®

 function on your device will automatically 

turn on and will be ready to connect with the «Microlife 

Connected » App after the device is switched on. 

Your device will automatically upload the data once it is 

connected to the smartphone.

11. Malfunctions and Actions to take

12. Cleaning and Disinfecting

Use an alcohol swab or cotton tissue moistened with 

alcohol (70% Isopropyl) to clean the silicone that touches 

the finger inside of the device. Also clean the finger being 

Descrip-

tion

Symptom/Possible 

causes

Solutions

SpO

2

 or 

pulse rate 

do not 

display 

normally.

1. Finger is not 

inserted 

correctly.

2. Patient SpO

2

 

value is too low to 

be measured.

3. There is exces-

sive illumination.

1. Retry inserting the 

finger.

2.  & 3. Measure more 

times. If you deter-

mine the product is 

working correctly, 

consult your doctor.

SpO

2

 or 

pulse rate 

is shown 

unstable.

1. Finger might not 

be inserted deep 

enough.

2. Excessive patient 

movement.

1. Retry inserting the 

finger.

2. Sit calmly and retry.

The 

device 

cannot be 

powered 

on.

1. No batteries or 

low battery 

power.

2. Batteries are not 

installed 

correctly.

3. The device may 

be damaged.

1. Replace the 

batteries.

2. Remove and rein-

stall the batteries.

3. Contact your local 

Microlife-Customer 

Service.

The 

display 

suddenly 

switches 

off.

1. The device is 

automatically 

powered off, 

when no signal 

was detected 

after 10±2 

seconds.

2. The battery 

power is too low 

to operate.

1. Normal.

2. Replace the 

batteries.

Descrip-

tion

Symptom/Possible 

causes

Solutions

Summary of Contents for OXY500 BT

Page 1: ...01 03 Distributor Microlife AG Espenstrasse 139 9443 Widnau Switzerland www microlife com Shenzhen Jumper Medical Equipment Co Ltd D Building No 71 Xintian Road Fuyong Street Baoan Shenzhen Guangdong...

Page 2: ...l Number Date of Purchase Specialist Dealer Microlife OXY 500 BT 5 4 3 2 1 9 10 6 7 8 9 V2 01 E Alm setup Alm on Beep off Demo off Restore ok Brightness 4 Exit V2 01 E Sounds Setup SpO2 Alm Hi 100 SpO...

Page 3: ...internet at www microlife com where you will find a wealth of invaluable information on our products Retain instructions in a safe place for future reference Stay healthy Microlife AG Table of Conten...

Page 4: ...r immerse this device in water or other liquids For cleaning please follow the instructions in the Cleaning and Disinfecting section Do not use this device if you think it is damaged or notice anythin...

Page 5: ...to know the oxygen saturation of a patient so that doctors can detect problems in a timely manner 4 Measurement Principles Principle of this fingertip pulse oximeter A mathemat ical formula is establi...

Page 6: ...turn on hold the button for about one second The device shows the settings menu AT Press or hold the ON OFF button to perform corresponding operations Hold it to set an item or press it to switch an o...

Page 7: ...once it is connected to the smartphone 11 Malfunctions and Actions to take 12 Cleaning and Disinfecting Use an alcohol swab or cotton tissue moistened with alcohol 70 Isopropyl to clean the silicone...

Page 8: ...guarantee service be required please contact the dealer from where the product was purchased or your local Microlife service You may contact your local Microlife service through our website www micro...

Page 9: ...ive maximum humidity Automatic switch off Automatically shut down in 10 2 seconds when no or low signal is detected Communication Bluetooth Low Energy 4 0 Compatibility iOS iOS 10 0 or newer Android A...

Page 10: ...8...

Reviews: