Microlife OXY500 BT Manual Download Page 6

4

The device must be able to measure the pulse 

properly to obtain an accurate SpO

2

 measure-

ment. Verify that nothing is hindering the pulse 

measurement before relying on the SpO

2

 

measurement.

The maximum application time at a single site 

should be less than 30 minutes

, in order to 

ensure correct sensor alignment and skin integ-

rity.

Inaccurate measurements may occur if:

The patient suffers from significant levels of dysfunc-

tional hemoglobin (such as carboxyhemoglobin or 

methemoglobin).

Intravascular dyes such as indocyanine green or meth-

ylene blue have been injected into the patient.

Used in the presence of high ambient light (e.g. direct 

sunlight). Shield the sensor area with a surgical towel if 

necessary.

There is excessive patient movement.

The patient experiences venous pulsations.

The patient has hypotension, severe vasoconstriction, 

severe anemia, or hypothermia.

The patient is in cardiac arrest or is in shock.

Fingernail polish or false fingernails are applied.

6. Inserting the batteries 

6

After you have unpacked your device, first insert the 

batteries. The battery compartment is on the bottom of the 

device. Remove the battery cover by sliding it in the direc-

tion shown. Insert the batteries (2 x 1.5 V, size AAA), 

thereby observing the indicated polarity.

Replace the batteries when the low power indi-

cator 

5

 appears on the display.

Always replace both batteries at the same time.

7. ON/OFF button 

1

 / Function button 

Press and release the ON/OFF button 

1

 to turn on, hold 

the button for about one second. The device shows the 

settings menu 

AT

. Press or hold the ON/OFF button to 

perform corresponding operations. Hold it to set an item, or 

press it to switch an option or switch the display mode. 

Press means no more than 0.5 seconds, while hold means 

more than 0.5 seconds.

Alert Sound Setting

Hold ON/OFF button 

1

 while the device is switched 

on.Settings menu (interface 1) is displayed 

AT

-a. Move «*» 

to the corresponding option, and hold the function button to 

set 

Alm

 to 

on

 and set 

Beep

 to 

off

. When 

Alm

 is set to 

on

 

and the measured values of the blood oxygen saturation 

and pulse rate go beyond the upper limit or lower limit, the 

device gives off an alert sound. When 

Beep

 is set to 

on

, a 

tick will be heard along with pulse beats during pulse rate 

measurement. While the «*» symbol stays on the 

Restore

 

option, hold the functional button to restore default settings.

Brightness Setting

Press ON/OFF button 

1

 to select the 

Brightness

 option 

and then hold the ON/OFF button to set the brightness to a 

value ranging from 1 to 5. The greater the value, the 

greater the brightness of the screen.

Alert Range Setting

On settings menu (interface 2) 

AT

-b, press the ON/OFF 

button 

1

 to switch between options. On this interface, you 

can set the upper limit and lower limit of 

SpO2 Alm

 and 

PR 

Alm

. While the «*» symbol stays on the +/- option, hold the 

functional button to set the option to + or -. 

In + mode, select the corresponding option and hold the 

ON/OFF button to increase the upper or lower limit; in - 

mode, hold the ON/OFF button to decrease the upper or 

lower limit. Move «*» to the 

Exit

 option, and hold the ON/

OFF button to return to the monitoring interface.

Summary of Contents for OXY500 BT

Page 1: ...01 03 Distributor Microlife AG Espenstrasse 139 9443 Widnau Switzerland www microlife com Shenzhen Jumper Medical Equipment Co Ltd D Building No 71 Xintian Road Fuyong Street Baoan Shenzhen Guangdong...

Page 2: ...l Number Date of Purchase Specialist Dealer Microlife OXY 500 BT 5 4 3 2 1 9 10 6 7 8 9 V2 01 E Alm setup Alm on Beep off Demo off Restore ok Brightness 4 Exit V2 01 E Sounds Setup SpO2 Alm Hi 100 SpO...

Page 3: ...internet at www microlife com where you will find a wealth of invaluable information on our products Retain instructions in a safe place for future reference Stay healthy Microlife AG Table of Conten...

Page 4: ...r immerse this device in water or other liquids For cleaning please follow the instructions in the Cleaning and Disinfecting section Do not use this device if you think it is damaged or notice anythin...

Page 5: ...to know the oxygen saturation of a patient so that doctors can detect problems in a timely manner 4 Measurement Principles Principle of this fingertip pulse oximeter A mathemat ical formula is establi...

Page 6: ...turn on hold the button for about one second The device shows the settings menu AT Press or hold the ON OFF button to perform corresponding operations Hold it to set an item or press it to switch an o...

Page 7: ...once it is connected to the smartphone 11 Malfunctions and Actions to take 12 Cleaning and Disinfecting Use an alcohol swab or cotton tissue moistened with alcohol 70 Isopropyl to clean the silicone...

Page 8: ...guarantee service be required please contact the dealer from where the product was purchased or your local Microlife service You may contact your local Microlife service through our website www micro...

Page 9: ...ive maximum humidity Automatic switch off Automatically shut down in 10 2 seconds when no or low signal is detected Communication Bluetooth Low Energy 4 0 Compatibility iOS iOS 10 0 or newer Android A...

Page 10: ...8...

Reviews: