9.

 

How to recall 12 readings in Memory Mode

This thermometer can recall the last 12 readings. To recall please  
follow sequence.

Pressing and releasing the START button after the last 12 readings 
have been recalled will resume the above sequence from reading 1.

10.  Cleaning and Storage

Use an alcohol swab or cotton tissue 
moistened with alcohol (70% Isopropyl) 
to clean the thermometer casing and the 
measuring probe. Ensure that no liquid 
enters the interior of the thermometer. 
Never use abrasive cleaning agents, thinners 
or benzene for cleaning and 

never immerse the device in water 

or other cleaning liquids.

 Take care not to scratch the surface of 

the probe lens and the display. Remove the battery from the device if 
it is not required for extended periods of time in order to avoid dama-
ge to the thermometer resulting from a leaking battery.

9

11.  Error messages

LCD Display

Display 
Meaning

Description

Recall Mode

Press the START button to 
enter Recall Mode when 
power is off.The memory 
icon 

«M»

 flashes.

Reading 1 
(the last rea-
ding)

Press and release the 
START button to recall 
the last reading. Display 1 
alone with memory icon.

Reading 2  
(the second 
latest reading)

Press and release the 
START button to recall the 
second latest reading.

Reading 12  
(the last  
reading)

Press and release the 
START button consecu-
tively to recall readings in 
succession, up to the last 
12 readings.

10

Display / Problem

Display 
Meaning

Possible cause and 
fault remedy

Measured 
Temperature too 
high

Displays 

«H»

 when 

measured temperature 
higher than 100.0 °C / 
212.0 °F.

Measured tempe-
rature too low

Displays 

«L»

 when 

measured temperature 
lower than 0 °C /  
32.0 °F.

Ambient tempera-
ture too high

Displays 

«H»

 in 

conjunction with the 
«

» when ambient 

temperature is higher 
than 40.0 °C /  
104.0 °F.

Ambient tempera-
ture too low

Display 

«L»

 in con-

junction with the «

» 

when ambient tempe-
rature is lower than  
10.0 °C / 50.0 °F.

Error function  
display

When system has 
malfunction.

Blank display

Please check if the 
battery has been 
loaded correctly. Also 
check polarity (<+> 
and <->) of battery.

Dead battery 
indication

If the steady battery 
icon is the only 
symbol shown on the 
display, the battery 
should be replaced 
immediately.

EN

Summary of Contents for IR 1DE1

Page 1: ...de fi vre Silent GlowTM technologie Koortsalarm Silent GlowTM technologie Illuminated Display Nite Glow Ecran r tro clair Verlicht Display 12 Data Memory 12 M moires 12 Geheugens Celsius Fahrenheit sw...

Page 2: ...is thermometer completely hygienic for use by the whole family Fever Alarm 10 short beeps and a red LCD backlight alert the patient that he she may have a temperature equal to or higher than 37 5 C Di...

Page 3: ...d for the purposes described in this booklet The manufacturer cannot be held liable for damage caused by incorrect application Always use the thermometer with a new undamaged Probe Cover for each meas...

Page 4: ...probe into the center part of the probe cover LCD Display Display Meaning Description All segments displayed Press the O I button to turn on the unit all segments will be shown for 2 seconds Memory T...

Page 5: ...mp ly turn OFF the unit press and hold the START button for 5 seconds until and the flashing C or F icon is shown on the LCD Switch the measurement scale between C and F by pressing the START button a...

Page 6: ...he memory icon M flashes Reading 1 the last rea ding Press and release the START button to recall the last reading Display 1 alone with memory icon Reading 2 the second latest reading Press and releas...

Page 7: ...Class IIa Technical alterations reserved 15 www microlife com Detailed user information about our thermometers and blood pressure monitors as well as services can be found at www microlife com 12 Repl...

Page 8: ...de risque de verre cass et d ingestion de mercure Totalement s r pour la prise de temp rature des enfants Embouts jetables rendant le thermom tre parfaitement hygi nique pour toute la famille Signal...

Page 9: ...nuel Le fabricant ne peut tre tenu pour responsable de dommages provoqu s par une application incorrecte A chaque prise utilisez toujours le thermom tre avec un nouvel embout jetable pour viter toute...

Page 10: ...reil introduire verticalement la sonde dans le centre transparent de l embout Affichage de l cran Signification de l affichage Description Tous les segments sont affich s Appuyer sur le bouton O I pou...

Page 11: ...nit de mesure une autre teindre l appareil appuyer sans rel cher pendant 5 secondes sur le bouton START Quand le bouton START est rel ch apr s ces 5 secondes l unit de mesure symbole C ou F va clignot...

Page 12: ...ignote Mesure 1 dernier r sultat Presser puis rel cher le bouton START pour appeler la derni re mesure Affichage 1 juste avec symbole m moire Mesure 2 le 2e r sultat le plus r cent Presser puis rel ch...

Page 13: ...taill e pour l utilisateur de nos thermom tres et de nos autotensiom tres ainsi que sur nos services est disponible sur www microlife fr 24 12 Remplacement de la pile Cet appareil est fourni avec une...

Page 14: ...ter volledig hygienisch door iedereen gebruikt kan worden Hoge temperatuurindicatie 10 korte piepsignalen en een rood LCD achtergrondverlichting maken de pati nt erop attent dat hij zij een temperatuu...

Page 15: ...gesteld voor schade veroorzaakt door onjuiste toepassing Om een infectie of besmetting te voorkomen dient u altijd de thermometer met een ongebruikt beschermkapje te gebruiken Alleen Microlife bescher...

Page 16: ...van de oorthermometer verticaal in het midden van het geplaatste beschermkapje LCD scherm Betekenis Omschrijving Alle segmenten verschijnen Druk op de O I toets om de thermometer te activeren alle sy...

Page 17: ...pnieuw op de START toets te drukken Als u van meeteenheid bent veranderd dient u 5 seconden te wach ten waarna het toestel automatisch naar de meetstand overschakelt 32 7 Gebruiksaanwijzingen Belangri...

Page 18: ...t apparaat uit staat Het geheugen symbool M knippert Resultaat 1 het laatste resultaat Druk op de START toets en laat deze los om het laatste resultaat te zien Scherm 1 alleen met geheugensymbool Resu...

Page 19: ...evice Directive 93 42 EEC klasse IIa Technische wijzigingen voorbehouden 15 www microlife nl Op de website www microlife nl vindt u gedetailleerde gebruiks informatie over onze thermometers bloeddrukm...

Reviews: