6
* Please immediately consult your doctor, if this or any other problem
occurs repeatedly.
If you think the results are unusual, please read through the
information in «Section 1.» carefully.
7. Safety, Care, Accuracy Test and Disposal
Safety and protection
Follow instructions for use. This document provides important
product operation and safety information regarding this device.
Please read this document thoroughly before using the device and
keep for future reference.
This device may only be used for the purposes described in these
instructions. The manufacturer cannot be held liable for damage
caused by incorrect application.
This device comprises sensitive components and must be treated
with caution. Observe the storage and operating conditions
described in the «Technical Specifications» section.
Protect it from:
- water and moisture
- extreme temperatures
- impact and dropping
- contamination and dust
- direct sunlight
- heat and cold
The cuffs are sensitive and must be handled with care.
Only pump up the cuff once fitted.
Do not use this device close to strong electromagnetic fields such
as mobile telephones or radio installations. Keep a minimum
distance of 3.3 m from such devices when using this device.
«
ERR 3
»
AO
-C
No pressure
in the cuff
An adequate pressure cannot be generated in
the cuff. A leak may have occurred. Check that
the cuff is correctly connected and is not too
loose. Replace the batteries if necessary.
Repeat the measurement.
«
ERR 5
» Abnormal
result
The measuring signals are inaccurate and no
result can therefore be displayed. Read through
the checklist for performing reliable measure-
ments and then repeat the measurement.*
«
HI
»
Pulse or cuff
pressure too
high
The pressure in the cuff is too high (over
299 mmHg) OR the pulse is too high (over
200 beats per minute). Relax for 5 minutes
and repeat the measurement.*
«
LO
»
Pulse too
low
The pulse is too low (less than 40 beats per
minute). Repeat the measurement.*
Error
Description Potential cause and remedy