11
BP W90
EN
invoice. Repair or replacement within guarantee does not prolong or
renew the guarantee period. The legal claims and rights of
consumers are not limited by this guarantee.
12. Technical Specifications
This device complies with the requirements of the Medical Device
Directive 93/42/EEC.
Technical alterations reserved.
Operating conditions:
10 - 40 °C / 50 - 104 °F
15 - 95 % relative maximum humidity
Storage conditions:
-20 - +55 °C / -4 - +131 °F
15 - 95 % relative maximum humidity
Weight:
140 g (including batteries)
Dimensions:
68 x 68 x 70 mm
Cuff size:
13.5 - 21.5 cm (5.25 - 8.5 inches)
Measuring procedure:
oscillometric, corresponding to Korotkoff
method: Phase I systolic, Phase V diastolic
Measurement range:
20 - 280 mmHg – blood pressure
40 - 200 beats per minute – pulse
Cuff pressure display
range:
0 - 299 mmHg
Resolution:
1 mmHg
Static accuracy:
pressure within ± 3 mmHg
Pulse accuracy:
± 5 % of the readout value
Voltage source:
2 x 1.5 V alkaline batteries; size AAA
Battery lifetime:
approx. 460 measurements
(using new batteries)
IP Class:
IP20
Reference to
standards:
EN 1060-1 /-3 /-4; IEC 60601-1;
IEC 60601-1-2 (EMC); IEC 60601-1-11
Expected service life:
Device: 5 years or 10000 measurements
Accessories: 2 years
Summary of Contents for BP W90
Page 87: ...85 BP W90 RU ...
Page 111: ...109 BP W90 HU ...
Page 125: ...123 BP W90 BG ...
Page 137: ...135 BP W90 RO ...
Page 149: ...147 BP W90 CZ ...
Page 161: ...159 BP W90 SK ...
Page 173: ...171 BP W90 TR ...
Page 187: ...185 BP W90 GR ...
Page 197: ...195 BP W90 AR ...