7
BP W150 IMT
EN
7. Battery Indicator and Battery change
Low battery
When the batteries are approximately ¾ empty the battery symbol
AO
will flash as soon as the instrument is switched on (partly filled
battery displayed). Although the instrument will continue to measure
reliably, you should obtain replacement batteries.
Flat battery – replacement
When the batteries are flat, the battery symbol
AO
will flash as soon
as the instrument is switched on (flat battery displayed). You cannot
take any further measurements and must replace the batteries.
Using rechargeable batteries
You can also operate this instrument using rechargeable batteries.
8. Error Messages
If an error occurs during the measurement, the measurement is inter-
rupted and an error message, e.g. «
ERR 3
», is displayed.
1. Open the battery compartment
3
by pulling off the cap.
2. Replace the batteries – ensure correct polarity as shown by the
symbols in the compartment.
3. To set date and time, follow the procedure described in «Section 2.».
)
The memory retains all values although date and time (and
possibly also set alarm times) must be reset – the year number
therefore flashes automatically after the batteries are replaced.
)
Use 2 new, long-life 1.5V, size AAA batteries.
)
Do not use batteries beyond their date of expiry.
)
Remove batteries if the instrument is not going to be used for
a prolonged period.
)
Only use «NiMH» type reusable batteries.
)
Batteries must be removed and recharged when the flat
battery symbol appears. They should not remain inside the
instrument as they may become damaged (total discharge as
a result of low use of the instrument, even when switched off).
)
Always remove the rechargeable batteries if you do not intend
to use the instrument for a week or more.
)
Batteries cannot be charged in the blood pressure monitor.
Recharge batteries in an external charger and observe the
information regarding charging, care and durability.
Error
Description Potential cause and remedy
«
ERR 1
» Signal too
weak
The pulse signals on the cuff are too weak.
Re-position the cuff and repeat the
measurement.*
Summary of Contents for BP W150 IMT
Page 13: ...11 BP W150 IMT EN...
Page 43: ...41 BP W150 IMT SV...
Page 63: ...61 BP W150 IMT DA...
Page 103: ...101 BP W150 IMT EE...
Page 105: ...103 BP W150 IMT RU 1 2 3 4 5 6 7 8 9 10 11 1 BF...
Page 106: ...104 o a e 15 4...
Page 108: ...106 3 1 2 3 4 5 6 13 5 21 5 5 25 8 5 7 8 1 9 10 11 AN 12 7 8 9 13 1...
Page 109: ...107 BP W150 IMT RU 4 AM 5 M 5 M AL M 17 17 M M M CL 30...
Page 110: ...108 M CL 6 AK 6 WHO 1 7 AO AO 1 3 2 3 2 2 1 5 AAA oc a NiMH...
Page 111: ...109 BP W150 IMT RU 8 ERR 3 ERR 1 ERR 2 ERR 3 ERR 5 HI 300 200 5 LO 40 1...
Page 112: ...110 9 2 Microlife...