Error messages/malfunctions
If an error occurs during a measurement, a long beep followed by two short beeps
is generated and the LCD displays the corresponding error code.
Possible cause
●
No pulse has been detected.
●
Unnatural pressure impulses
influence the measurement
result.
Reason:The arm was moved
during the measurement.
●
The inflation of the cuff takes too
long. The cuff is not correctly
seated.
●
The difference between systolic
pressure and diastolic pressure is
too far away from acceptable
and reasonable.
●
Due to unstable conditions
during measurements, it is not
ssible to calculate an average
result.
●
Low battery
Error
Err 1
Err 2
Err 3
Err 5
Err 6
4.4
Remedy
●
Ensure that the cuff is being worn
correctly,and that you have your
arm at the heart level.
●
Avoid unnecessary movement or
talking.
●
Ensure that the cuff is being worn
correctly.
●
Ensure that the cuff is being worn
correctly and that you have been
inactive for a sufficient time before
marking the measurement.
●
Avoid unnecessary movement and
talking.
●
Replace batteries.