selecting the correct cuff
aponorm
®
by microlife offers different cuff sizes. Select the cuff
size to match the circumference of your upper arms (measured
by close fitting in the centre of the upper arm).
AfiB/MAM mode (highly recommended)
In AFIB/MAM mode, 3 measurements are automatically taken in
succession and the result is then automatically analysed and
displayed. Because blood pressure constantly fluctuates, a
result determined in this way is more reliable than one produced
by a single measurement. AFIB detection is only activated in
AFIB/MAM mode.
• After pressing the START/STOP button
1
, the selected
AFIB/MAM mode appears in the display as the MAM-
symbol
26
.
• The bottom, right hand section of the display shows a 1, 2
or 3 to indicate which of the 3 measurements is currently
being taken.
• There is a break of 15 seconds between the
measurements (15 seconds are adequate according to
«Blood Pressure Monitoring, 2001, 6:145-147» for
oscillmetric instruments). A count down indicates the
remaining time.
• The individual results are not displayed. Your blood
pressure will only be displayed after all 3 measurements
are taken.
• Do not remove the cuff between measurements.
• If one of the individual measurements was questionable,
a fourth one is automatically taken.
Cuff size
for circumference of upper arm
S
17 - 22 cm
M
22 - 32 cm
M - L
22 - 42 cm
L - XL
32 - 52 cm
Only use
aponorm
®
by microlife cuffs.
Connect the cuff to the device by inserting the cuff
connector
7
into the cuff socket
3
as far as it will go.
selecting the user
This device allows to store the results for 2 individual users.
Before each measurement
, set the user switch
9
for the
intended user: user 1 or user 2.
User 1: slide the user switch
9
upwards to the user 1 icon.
User 2: slide the user switch
9
downwards to the user
2 icon.
The first person to measure should select user 1.
select the measuring mode: standard or AfiB/MAM mode
This device enables you to select either standard (standard
single measurement) or AFIB/MAM mode (automatic triple
measurement).
To select standard mode, slide the AFIB/MAM switch
8
on
the side of the device downwards to position «1» and to select
AFIB/MAM mode, slide this switch upwards to position «3».