08
Loading and Unloading
Operating Instructions
Loading and unloading the PDQeX
After running the device, the DNA will be in
the collection tubes.
First remove the PDQeX extractor
cartridges and dispose. Slide out the
drawer.
Remove and cap the collection tubes and
store at -20
°
C.
The door of the device will lock while the block is hot.
For sequential runs, make sure the block temperature has fallen below the first temperature used in your
program.
Carefully remove tubes to prevent contamination of the block.
Tubes cannot be re-used if you abort a run.
Notes