15
Cleaning, Sterilization, Inspection, and Maintenance
Cleaning, Sterilization, Inspection, and Maintenance
Collectiontraysandcartridgeholdersshouldbecleanedandsterilizedinaccordancewithlaboratoryprocedure.
The UV treatment option on the control screen gives a 30-minute UV light exposure (see pages 10-11).
In the event of gross contamination of the heating block or device, clean with a cotton swab and wipe with 0.5%
bleach followed by a thorough removal of the bleach with water-dampened swabs and wipes.
Aluminum parts of this device should not be soaked in virkon or bleach. Extended use of chemical agents such
as these will corrode the parts.
The following label may be displayed on the device:
The PDQeX has been designed to require no operator maintenance.
Cleaning and Sterilization
WARNING: Do not get liquids inside the device.
If bleach must be used, be sure ALL of it is removed.
Safety Label
Hazard Symbol
English
Francais
UV
WARNING: Avoid looking
directly at the UV light
AVERTISSEMENT: Évitez
de regarder directement à
la lumière UV
Inspection and Maintenance
Disposal of Waste
Used PDQeX cartridge tubes should be disposed as hazardous waste in accordance with laboratory protocols.