The system uses standard path lengths to measure the sample’s optical density on the
pedestal or through cuvette. In case of an OD value increase due to various issues,
the OD values can be adjusted. It is not recommended for users to make such changes,
but they are nonetheless available when such functionality is necessary. OD values can be
adjusted as below.
2. Touch the Calibration button at the left bottom corner to open the Calibration Panel.
(Users must be logged in as Master in order to gain access to the Calibration Panel.)
3. Microvolume OD values can be adjusted in the
box under Nano, and Cuvette OD values can
be adjusted in the box under Cuvette.
4. These values can be adjusted from -1000 to
1000 and adjustment of 1 unit corresponds to
OD value of 0.001.
5. For instance, in order to increase the OD value
by 0.01, type in 10 into the value field and to
decrease the OD value, by 0.05, type in -5 into
the value field and touch Okay.
3. User Interface
33
1. Open the Setup panel by touching the Setup button from the Main page.
Setup Panel
Calibration Panel