Kinetic Measurement is to perform multiple spectrometry measurements on samples
in accordance with users’ wavelength and time lengths.
Title
Title of the measurement.
Wavelength (nm)
Wavelength of the measurement between 200
– 1,100 nm.
Measurement Time (min)
Overall measurement.
Interval Time (sec)
Interval time between detections.
3. User Interface
Procedure
1. Proceed to Kinetic page from the Main page.
2. Insert the title of the measurement.
3. Set the Wavelength, Measurement Time and Interval Time. Users may calculate the
amount of the detection by dividing Measurement Time with Interval Time, and round up
the number. (e.g. if the Measurement Time is set to 1 min, and Interval Time is set to 9
seconds, then there will be total 6 detections.)
4. Set a blank value for the measurement. (Check pg. 8)
5.
Insert the target sample in a cuvette. (Minimum Volume: 1mℓ)
6. Touch Read button to start the measurement.
7. Check the results on the screen. The resulting data will be saved automatically,
and can be searched in the search panel.
Kinetic Page
22