Kinetic Measurement is to perform multiple spectrometry measurements on samples 

in accordance with users’ wavelength and time lengths.

Title

Title of the measurement.

Wavelength (nm)

Wavelength of the measurement between 200 

– 1,100 nm.

Measurement Time (min)

Overall measurement.

Interval Time (sec)

Interval time between detections.

3. User Interface

Procedure

1. Proceed to Kinetic page from the Main page.

2. Insert the title of the measurement.

3. Set the Wavelength, Measurement Time and Interval Time. Users may calculate the 

amount of the detection by dividing Measurement Time with Interval Time, and round up 

the number. (e.g. if the Measurement Time is set to 1 min, and Interval Time is set to 9 

seconds, then there will be total 6 detections.)

4. Set a blank value for the measurement. (Check pg. 8)

5.

Insert the target sample in a cuvette. (Minimum Volume: 1mℓ) 

6. Touch Read button to start the measurement.

7. Check the results on the screen. The resulting data will be saved automatically,

and can be searched in the search panel.

Kinetic Page

22

Summary of Contents for Nabi

Page 1: ...Nabi User Manual www md best com All Rights Reserved User Manual Ver 3 6...

Page 2: ...www md best com All Rights Reserved Nabi User Manual 1...

Page 3: ...INTRODUCTION 4 1 1 System Package 5 1 2 Main Instrument 5 1 3 Specification 6 2 BASIC USE 8 2 1 Pedestal 8 2 2 Cuvette 9 2 3 Blank Setting 9 2 4 Keyboard 10 3 USER INTERFACE 11 3 1 Log In 11 3 2 Main...

Page 4: ...Curve 17 3 5 2 Bradford Lowry BCA Biuret and Others 19 3 5 3 Direct UV 20 3 6 End Point and OD600 21 3 7 Kinetic 22 3 8 Spectrum 23 3 9 Search 24 3 9 1 Overlay in Search 26 3 9 2 Data Transfer 28 3 1...

Page 5: ...e sample with high accuracy outstanding precision and superb repeatability The system can be upgraded with ease via USB Flash Drive EASY UPGRADES Samples in a universal cuvette or microscale volume ca...

Page 6: ...kage includes Main Instrument Pedestal Cover Cuvette Stick Cuvettes Power Cord User Manual Lower Pedestal Cuvette Port Upper Pedestal USB Port 1 USB Port 2 LCD Touch Screen Power Switch Power Cord Por...

Page 7: ...11 0 inch H 8 6 inch Weight 3 kg Operator Voltage 100 240V 50 60 Hz Power Consumption 80 100 W Warranty 1 Year LCD Touch Panel 7 0 inch General Specification Absorbance Precision 1 at 100 ng l Absorba...

Page 8: ...NOT try to disassemble the instrument in any situation Use an electric supply that conforms to industry standards Place the instrument in a clean environment and away from other devices that vibrate...

Page 9: ...se of the Pedestal The Pedestal has upper and lower pedestals Targeted microscale sample should be gently pipetted on the center top of the lower pedestal and the Cuvette Stick should be inserted duri...

Page 10: ...cuvette and the cuvette should be gently inserted in the Cuvette Port for the measurement The Pedestal Cover should be covered during the cuvette measurement The direction of the light pathway during...

Page 11: ...ument The keyboard will automatically pop up when any letter box e g title is touched When button located at the bottom left corner of the keyboard is touched it will convert to the number and special...

Page 12: ...nt with specified ID Guest Logs in as a guest When the instrument is turned on the Log In page will appear as above Users can enter their ID and Password and touch the Log In button to log in or can l...

Page 13: ...surement page pg 13 PROTEIN Opens Protein Menu pg 15 OD600 Opens OD600 Measurement page pg 19 END POINT Opens End Point Measurement page pg 19 KINETIC Opens Kinetic Measurement page pg 20 SPECTRUM Ope...

Page 14: ...rive connected Refer to pg 13 BLANK Measures and sets the blank value READ Performs measurement according to the settings 13 3 User Interface Users may capture a screenshot of the screen by touching t...

Page 15: ...NA samples Oligo Measures Oligonucleotide samples Overlay Button Goes to Overlay page Title Title of the measurement Conc Shows the concentration of the sample 260nm 10mm path Shows the Abs Value at 2...

Page 16: ...eck the results on the screen The resulting data will be saved automatically and can be searched in the search panel Nucleic Acids Overlay Measurement Overlay page is to compare up to 5 results Overla...

Page 17: ...simply retouch the Protein button or touch Nabi logo banner BRADFORD Opens Bradford Measurement page pg 17 LOWRY Opens Lowry Measurement page pg 17 BCA Opens BCA Measurement page pg 17 BIURET Opens B...

Page 18: ...e Save Open and Reset buttons Measurement Loads Measurement page It is a default page when the users enter any pages under Protein Standard Curve Loads Standard Curve page for a Standard Curve setting...

Page 19: ...th different concentration to setup the Standard Curve 2 Set the blank value Check pg 8 3 Insert the target sample in a cuvette Minimum Volume 1m 4 Touch Read button to measure 5 When the measurement...

Page 20: ...e results on the screen The resulting data will be saved automatically and can be searched in the search panel 19 3 User Interface Bradford Page Biuret Page BCA Page Lowry Page Others Page There are v...

Page 21: ...target sample on the Lower Pedestal and close the Head 6 Touch Read button to start the measurement 7 Check the results on the screen The resulting data will be saved automatically and can be searched...

Page 22: ...f the measurements End Point and OD600 Measurement is to perform spectrometry measurement on Cell Culture samples and etc Procedure 1 Proceed to End Point or OD600 page from the Main page 2 Insert the...

Page 23: ...Main page 2 Insert the title of the measurement 3 Set the Wavelength Measurement Time and Interval Time Users may calculate the amount of the detection by dividing Measurement Time with Interval Time...

Page 24: ...avelength Procedure 1 Proceed to Spectrum page from the Main page 2 Insert the title of the measurement 3 Set the range of Wavelength 4 Set a blank value for the measurement Check pg 8 5 Insert the ta...

Page 25: ...the search Title Search Finds data under the title by matching keywords Date Time Finds data under specified dates and times Users can load a calendar by touching From and To buttons and select the d...

Page 26: ...ch data edit and delete the selected data and transfer the data to a USB Flash Drive in FAT format In DNA RNA Users can delete more than one datum at once by selecting multiple data Selected data are...

Page 27: ...Data Transfer Transfers the displayed list to USB Flash Drive The USB Flash Drive must be in FAT Format Back Goes back to previous page Overlay Panel Overlay Panel is to compare the results up to 5 It...

Page 28: ...ransferred files will be saved in the Nabi folder Please refer to the diagram on pg 28 for detailed location of the files Nucleic Acids Kinetic Spectrum and Direct UV will include Raw Data for each me...

Page 29: ...file will be titled as the name inserted during the process Folder Name User ID Nabi ID 1 ID 2 Nucleic Acids Kinetic Protein End Point Spectrum Raw Data Raw Data Raw Data Direct UV Raw Data OD600 dsDN...

Page 30: ...te the selected account in the Account page Master account cannot be deleted and its password can be changed in the Setup page Current ID Shows ID of current account logged in Swap Swaps between the c...

Page 31: ...e User Interface after using the instrument and when it is completely shut down turn off the Power Switch at the back of the instrument When the power is forcibly turned off by the power switch withou...

Page 32: ...recommended for users to access the page and change features without manufacturer s supervision Time Setup Changes the time in the system Update Updates the interface Format Database Deletes all the d...

Page 33: ...minute DO NOT Turn off the instrument or try other functions in the user interface during the update 6 The User Interface will be automatically restart to the new version 1 Download the update file M...

Page 34: ...bottom corner to open the Calibration Panel Users must be logged in as Master in order to gain access to the Calibration Panel 3 Microvolume OD values can be adjusted in the box under Nano and Cuvett...

Page 35: ...k The eyes will protect you from bad spirits while it is on Setup Panel 1 Open the Setup panel by touching the Setup button from the Main page 2 Touch the Calibration button at the left bottom corner...

Page 36: ...MicroDigital Co Ltd 7th floor CS Building 15Pangyo ro 228 beon gil BunDang Gu SungNam Si GyungGi Do 13487 Korea Tel 82 31 701 2225 Fax 82 31 702 2225 info md best com www MD Best com...

Reviews: