6
Integrate the meter with the control
system
Topics covered in this chapter:
•
•
•
•
•
Configure HART/Bell 202 communications
•
Configure Modbus communications
•
Digital Communications Fault Action
6.1
Configure Channel B
Display
OFF-LINE MAINT
>
OFF-LINE CONFG
>
IO
>
CONFIG CH B
ProLink III
Device Tools
>
Configuration
>
I/O
>
Channels
Field Communicator
Configure
>
Manual Setup
>
Inputs/Outputs
>
Channels
>
Channel B
Overview
Depending on your device, you can configure Channel B to operate as either an mA output
or a discrete output.
Restriction
You cannot configure Channel B on the CDM TPS device. On this device, Channel B always operates
as a TPS output.
Prerequisites
The configuration of Channel B must match the wiring. See the installation manual for your
device.
To avoid causing process errors:
•
Configure Channel B before configuring the mA output or discrete output.
•
Before changing the channel configuration, ensure that all control loops affected by
the channel are under manual control.
Procedure
Set Channel B as desired.
Integrate the meter with the control system
Configuration and Use Manual
83
Summary of Contents for CDM100M
Page 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Page 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Page 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Page 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Page 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Page 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Page 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Page 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Page 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...