Troubleshooting
Top-Line Video Endoscope
7.2
Video Display are not Working Properly
There is nothing to see on the LCD screen: The
battery is almost empty.
Recharge the battery.
Still images or videos cannot be recorded: There
is no SD card in the device.
Insert the SD card into the video endoscope.
Off: The battery is not charged enough.
Charge the battery.
None of the functions work or hang.
Switch off the video endoscope and restart the
device.
If the functions cannot be restored:
Reset
Remove the cover from the battery compart-
ment. Remove the battery to reveal the reset
button.
Insert a toothpick or another non-conducting
tool into the reset hole and push the reset but-
ton to perform a system reset.
7.3
Articulation is Difficult
Locking may be fully or partially engaged.
Release locking.
Articulation cables may be constricted by bends in
the flexible probe.
Keep the flexible probe as straight as possible
during inspection.
The probe tip may be blocked because of their
position within the device.
Reposition the probe tip by rotating the flexible
probe or moving the probe tip forwards and
backwards.
Do not apply excessive force on control levers.
> Damage or destruction of the probe tip
NOTICE