Page 32
Operation and Setting
optoCONTROL CLS-K
6.2.5.2
Linearization
1
Based on the basic function the linear part of the function is prepared with P3 “zero point“ and P2
“steepness“.
2
This adjustment requires repetitions as the settings of the potentiometer interfere each other.
3
A multimeter 0 – 10 VDC is connected to the measuring points “GND” and “U”
(S2: “2“; 10 V = 20 mA) for the adjustment procedure.
4
Start value for zero point approximately – 2 V.