background image

 

ILD1700 Tool  

 

___________________________________________________________________________ 
Micro-Epsilon 

 

Page 13 

 

5.1.2  Y-Axis Scaling Properties 

This mode automatically adjusts the y-scale of the display to zoom in on the 
measured values as close as possible. 
In manual mode the value range can be set by entering the min. and max. 
limits of the y-axis. 

5.1.2.1  Flip Y axis 

Turns the measured signal upside down. Can be used for example to display 
the surface of a measurement target.  

5.1.2.2  Y-Scaling Type 

The scaling of the y-axis can be set to one of the following methods. 

• 

Manual Y range: 
Enables the edit fields for the user can enter the upper and lower limit 
of the y–axis. 

• 

Autoscale Y range: 
The program searches the smallest and greatest value in the visible x-
range and sets the y-limits to those values. 

• 

Y center on data: 
This scaling mode requires the parameter 

Range(mm

), here called 

Y_range. 
The program calculates the average of the measurement values in 
the visible x-range. Then it sets the upper limit of the y-axis to 
Y_range and the lower limit to average- Y_range. 

 

5.1.2.3  Range (mm) 

This parameter determines the area of interest around the mean value of the 
measurement signal. 

5.2  Data Acquisition 

The purpose of this page is to acquire data, only. It does not change any of 
the sensor’s parameters. There is one exception: it sets the transfer mode 
from ASCII to binary. 
This data acquisition module requires a baud rate of 115200. Also the laser 
diode must be turned on. 

5.2.1  Measurement Properties 

5.2.1.1  Data Acquisition 

The button shows ‘

Start DAQ

’ if no connection to a sensor was established. To 

connect to a sensor and to start the measurement press the button once. If 
the connection was successful, the button shows ‘

Stop DAQ

’. Another click on 

the button stops the measurement. 

5.2.1.2  Hold last valid value 

If no valid target can be recognized the last valid value is kept and put out, 
otherwise an error value is put out to indicate that the sensor has ‘lost’ its 
target. 

5.2.2  AutoZero Properties 

This section offers a software distance calibration of the sensor position. 

Summary of Contents for ILD1700

Page 1: ...User manual for the ILD1700 Tool Version 1 4...

Page 2: ...Menu bar 6 3 4 1 FILE 6 3 4 2 PROGRAM 6 3 4 3 LANGUAGE 6 3 4 4 6 4 SENSOR CONFIGURATION 7 4 1 Configuration 7 4 1 1 CONNECTION SETTING 7 4 1 2 SENSOR PARAMETERS 8 4 1 3 CALIBRATION 8 4 1 4 ADDITIONAL...

Page 3: ...__________________ Micro Epsilon Page 3 5 3 Data Storage 14 5 3 1 DATA EXPORT PROPERTIES 14 5 4 Distance View 14 5 4 1 DISTANCE 14 5 4 2 STATISTICS 14 5 4 3 LIMIT OUTPUT PROPERTIES 14 5 4 4 FREEZE MOD...

Page 4: ...lution of 1024x768 Sensor controller firmware version 1 7 or higher IEEE422 RS422 to RS232 converter USB to Serial Port driver version 1 00 2154 FTSER2K SYS or higher 2 2 Software installation Start t...

Page 5: ...priate regional settings only Fig 2 Main Menu 3 1 Show Sensor Manual If you have a PDF viewer installed on your computer the documentation of the ILD1700 Sensor Controller is shown Since the documenta...

Page 6: ...4 2 2 Data Acquisition Switches to the data acquisition and storage program 3 4 2 3 Main Menu Shows the main page again 3 4 3 Language Determines the language of the descriptions and the fonts of the...

Page 7: ...D1700 Sensor Configuration 4 1 1 Connection Setting The COM port the ILD1700 controller is connected to Supported port numbers are COM1 to COM16 The sensor uses a RS422 to USB interface that is mapped...

Page 8: ...see Sensor Information The sensor manual contains a table that shows more on that 4 1 2 3 Output Type The mode RS422 activates the digital data output The items 4 20mA and 0 10V activate the analog da...

Page 9: ...binary mode when started The ASCII mode may be more suitable when using the HyperTerminal or other diagnostic tools to communicate with the sensor 4 1 4 3 Laser Turns the laser diode on or off 4 1 4 4...

Page 10: ...on the parameter Limit Assignment 4 1 6 Sensor Information Displays some general information about the sensor 4 1 6 1 Sensor Range Measurement range of the sensor 4 1 6 2 Read Sensor Info This button...

Page 11: ...___ Micro Epsilon Page 11 4 1 9 Sensor Fallback Reset To Factory Settings Overwrites the parameters stored in the sensor with the default parameters For example sets the baud rate to 115200 Reboot Sen...

Page 12: ...200 8 n 1 5 1 1 1 COM Number of the COM port the ILD1700 sensor is connected to see 4 1 1 5 1 1 2 Block size Determines the number of values to read at once A smaller block size reduces the reaction t...

Page 13: ...ment values in the visible x range Then it sets the upper limit of the y axis to average Y_range and the lower limit to average Y_range 5 1 2 3 Range mm This parameter determines the area of interest...

Page 14: ...ues are displayed in the plot graph because only the displayed values are stored when using the button Store displayed samples 5 4 Distance View This view displays the currently acquired or buffered m...

Page 15: ...ower hysteresis If the measurement value rises above this threshold the lower LED is set to green This threshold is displayed in orange 5 4 3 6 Lower limit If the measurement value falls below this li...

Page 16: ...uffers and switch into live mode 5 4 5 F1 F2 Upper limit This LED displays the state of the output for the upper limit A red LED indicates the upper limit is violated and the corresponding limit outpu...

Page 17: ...ILD1700 Tool ___________________________________________________________________________ Micro Epsilon Page 17...

Reviews: