Page 106
Appendix| ASCII Communication with Sensor
optoNCDT 1420
A 3.3.13
Key Function
A 3.3.13.1 KEYFUNC, Choose Key Function
KEYFUNC NONE | MASTER | TEACH
Choice of key function.
- NONE: key has no function
- MASTER: key is used for mastering
- TEACH: key is used for teaching
A 3.3.13.2 KEYLOCK, Set Keylock
KEYLOCK NONE|ACTIVE|AUTO <time>
Choice of keylock.
- NONE: key works permanently, no keylock
- ACTIVE: keylock is activated right after reboot
- AUTO: keylock is only activated <time> minutes after reboot
A 3.4
Measurement
A 3.4.1
TARGETMODE, Measurement Mode
TARGETMODE STANDARD|MULTISURFACE|PENETRATION
1
Choice of material dependent presets, see Chap.
.
- STANDARD: suitable for materials, e.g. made of ceramics, metal, plastics or wood
- MULTISURFACE: suitable for materials with changing surfaces, e.g. PCB or hybrid materials
- PENETRATION: suitable for materials with strong penetration depth of the laser light
A 3.4.2
MEASPEAK, Choice of the Peak in the Video Signal
MEASPEAK DISTA|DIST1|DISTL
- DISTA: output of peak with highest amplitude (standard)
- DIST1: output of first peak
- DISTL: output of last peak
A 3.4.3
MEASRATE, Measuring Rate
MEASRATE 0.25|0.5|1|2|4
Choice of measuring rate in kHz.