MDM50 HP Version User’s Manual
14
97478 Issue 1.4, October 2019
MAINTENANCE
5.6
Troubleshooting
Symptom
Cause
Actions
Display shows “
ErrI
”
Transmitter disconnected
or sensor element /
sensor thermistor fault.
NOTE: This error is
normally displayed
during the first few
seconds while the
instrument is starting
up.
Check transmitter cable is securely
connected inside instrument.
Check instrument electronics with
4-20mA source.
Exchange transmitter.
Display shows “
ErrL
”
Transmitter reading
under-range.
Transmitter current
signal
< 4mA.
Transmitter element
open circuit.
Check instrument electronics with
4-20mA source.
Exchange transmitter.
Display shows “
ErrH
”
Transmitter reading
over-range.
Transmitter current
signal > 20mA.
Sensor element short
circuit.
Sensor may have been exposed
to saturation conditions or liquid
contamination. Check that filter and
sensor guard are clean and dry.
If liquid water has contacted sensor,
or sensor block, then disassemble
and dry thoroughly. After drying, it
is recommend to purge assembled
instrument with very dry (-75°Cdp
(-103°Fdp) / 1ppm
V
moisture) air for
12 to 24 hours.
Exposure to other contaminants can
cause lasting damage and may require
transmitter to be exchanged.
Check instrument electronics with
4-20mA source.
Display flickers on
and off
Battery voltage low.
Connect charger.