c)
Measuring voltage ≤ 600V, uses terminal 3 and 1;
d)
Measuring voltage
>
600V, uses terminal 3 and 4;
3. Basic testing method for accuracy
(1) According to the terminal definition table to connect the testing circuit;
(2) The testing must to be done by under the following conditions
;
Auxiliary power: 24V±1
%
, ripple≤5mV
Ambient temperature: 25
℃
±2
℃
Relative humidity: RH(45
~
75)
%
Accuracy for Signal Power Source instrument: 0.05
(3) warming up the transducer for 3 minutes
(4) using output monitoring meter V1 to measuring the output of signal power source
instrument, set a any input value which is within the measuring range of the
transducer Vr. (for e.g. if input value is 100V, the output is 0-5V, the expected output
value Vz should be calculated as:
Vz=
(
5V-0V
)
× Vr/100V
(5) using output monitoring meter V2 to measuring the output voltage V0 of the
transducer, the basic introduced errorγof the transducer should be calculated as:
γ= (Vo
-
Vz)/
(
5V-0V
)
×100
(6) Repeating (4) (5), if calculated absolute value is less than the given accuracy value
of the transducer; it shows the transducer’s accurate grade is qualified.
3
Transducer
U
X
4
Resistor