(3) Warming up the transducer for 3 minutes

(4) Using output monitoring meter V1 to measuring the output of signal power

source instrument, set any input value which is within the measuring range of

the transducer Vr.(e.g. if input range is 100V, the output is 0-5V ). the expected

output value Vz should be calculated as:

Vz=5V×Vr/100V

(5) Using output monitoring meter V2 to measuring the output voltage V0 of the

transducer, the basic introduced error

γ

of the transducer should be calculated

as:

γ= (Vo-Vz)/5V×100

(6)

Repeating step (4) (5), if calculated absolute value is less than the given

accuracy value of the transducer, it shows the transducer’s accurate grade is

qualified.

Caution:

Caution:

Caution:

Caution:

1.

Pay attention to the auxiliary power information, especially the auxiliary power grade,

and polarity, other wise will damage the product.

2.

Pay attention to the wire connection; wrong terminal connection will cause malfunction

of the product and even damage the product;

3.

Don’t dismantle the product, and carry with care to avoiding bump and fall of the

product;

4.

If the product has been using under the environment with strong magnetic field

interference, please pay attention to the shield of input wire, and the output signal wire

should be as short as possible. For product intensive installation, the space between

Power
Source

Transducer

Auxiliary power

V

1

V

2

5

3

4

6

1

3

+

+

-

Summary of Contents for WBV342S01

Page 1: ...ltage Voltage Voltage Voltage Transducer Transducer Transducer Transducer www wb my com wblch wbdz cn Technical Service 86 816 2278271 Quality Complain 86 816 2278273 Fax Line 86 816 2281934 ISO9001 ISO14000 GB T2800 CE Certified Quality Warranty Any quality problem found in WB series products we offer Three years free charge of repair the products and six months guaranteed free charge of change a...

Page 2: ...ssioning and on careful operation maintenance and servicing The operating manual for the equipment gives instructions for its installation commissioning and operation However the manual cannot cover all conceivable circumstances or include detailed information on all topics In the event of questions or specific problem do not take any action without proper authorization Contact the appropriate WB ...

Page 3: ...2001 Product Product Product Product D D D Description escription escription escription and and and and Application Application Application Application WBV342S01 converts DC input voltage into a load independent output signal DC voltage 0 5V or 0 10V It has adopted modulation demodulation isolation principle and average measurement method for real time measurement of DC voltage from electric net T...

Page 4: ...DC 0 5 10 V 3 Accuracy class 0 2 4 Input Impedance R when the input voltage Vx 1V Ri Vx 10kΩ V when the input voltage Vx 1V Ri 1MΩ 5 Linear Range 0 120 of nominal input 6 Responding Time 150ms 7 Over Load Capacity 2 times of nominal input continuous 8 Output current 5mA 9 Auxiliary Power 12Vdc 24Vdc 10 Current consumption for static status 10mA 11 Isolation Voltage 2500Vdc for 1 minute 12 Output R...

Page 5: ...lation 2 The product has calibrated before out of factory After correctly wiring it can be powered and used immediately But for further precise signal sampling user needs to warm up the product for 3 minutes before use a The auxiliary power requires isolation voltage 2000VAC DC current output ripple 10mV and multiple converters can share with one set auxiliary power together b When measuring dc vo...

Page 6: ... 100 6 Repeating step 4 5 if calculated absolute value is less than the given accuracy value of the transducer it shows the transducer s accurate grade is qualified Caution Caution Caution Caution 1 Pay attention to the auxiliary power information especially the auxiliary power grade and polarity other wise will damage the product 2 Pay attention to the wire connection wrong terminal connection wi...

Page 7: ...There is no lightening strike prevention circuit design in this product For out door and hazardous environment using please add protective alternatives 7 This product uses fire prevent ABS crust its temperature withstand is only limited as 85 higher than this limitation will cause the product deformation Please use and store carefully 8 This product can t be used to measure the pulse DC voltage si...

Reviews: