Auxiliary power:
24V
±
1
%,
ripple
≤
5mV
Ambient temperature:
25
℃±
2
℃
Relative humidity
: RH(45
~
75)
%
Accuracy for Signal Power Source instrument:
0.05
(3) Warming up the transducer for 3 minutes
(4) Based on the indication of 3-phase AC Calibration Meter, set one set of
any
input value which is within the measuring range of the transducer Ur (line
voltage), Ir, COSφ, and the 3-phase input is balanced. (e.g. if input value is
100V and 5A, the output is 4-20mA, the expected output value Iz should be
calculated as:
Iz =
(
20mA
-
4mA)×│(Ur×Ir×COSφ)/(100V×5A)│
+
4mA)
)
(5) Using output monitoring meter V2 to measuring the output voltage V0 of the
transducer, the basic introduced error
γ
of the transducer should be calculated
as:
γ= (Vo-Iz×250Ω)/[(20mA-4mA)×250Ω]×100
Repeating step (4) (5), if calculated absolute value is less than the given accuracy value of the
transducer, it shows the transducer’s accurate grade is qualified.
Caution:
Caution:
Caution:
Caution:
1.
Pay attention to the auxiliary power information, especially the auxiliary power grade,
and polarity, other wise will damage the product.
2.
Pay attention to the wire connection; wrong terminal connection will cause
malfunction of the product and even damage the product;