background image

Table 204 

 

 
 
 
 
 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Guidance and manufacturer’s declaration – electromagnetic immunity 

The AES-U181 is intended for use in the electromagnetic environment specified below. 

The customer or the user of the AES-U181 should assure that it is used in such an environment. 

Immunity 
test 

IEC 60601 
test level 

Compliance 
level 

Electromagnetic environment –guidance 

Conducted 
RF 
IEC 61000-
4-6 
Radiated 
RF 
IEC 61000-
4-3 

3 Vrms 
150 kHz to 
80 MHz 
3 V/m 
80 MHz to 
2,5 GHz 

Not 
applicable 
3 V/m 

Portable and mobile RF communications equipment 
should be used no closer to any part of the AES-U181, 
including cables, than the recommended separation 
distance calculated from the equation applicable to the 
frequency of the transmitter. 
Recommended separation distance 
d=1,2 

d=1,2 

p 80 MHz to 800 MHz 

d=2,3 

p 800 MHz to 2,5 GHz 

where P is the maximum output power rating of the 
transmitter in watts (W) according to the transmitter 
manufacturer and d is the recommended separation 
distance in meters (m). 
Field strengths from fixed RF transmitters, 
as determined by an electromagnetic site 
survey, a should be less than the compliance level in 
each frequency range. b 
Interference may occur in the vicinity of 
equipment marked with the following symbol:   

 

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. 
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is 
affected by absorption and reflection from structures, objects and people. 

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) 
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV 
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic 
environment due to fixed RF transmitters, an electromagnetic site survey should be considered. 
If the measured field strength in the location in which the AES-U181 is used exceeds the 
applicable RF compliance level above, the AES-U181 should be observed to verify normal 
operation. If abnormal performance is observed, additional measures may be necessary, such 
as reorienting or relocating the AES-U181. 

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. 

Summary of Contents for AES-U181

Page 1: ...ype blood pressure Monitor AES U181 The monitor features blood pressure measurement pulse rate measurement and auto save the result The design provides you with many years of reliable service Reading taken by the AES U181 is equivalent to those obtained by a trained observer using the cuff and stethoscope auscultation method This manual contains important safety and care information and provides s...

Page 2: ... applied part Caution Consult accompanying documents Do not dispose this product as unsorted municipal waste Collection of such waste separately for special treatment is necessary Stand by button Transport package shall be kept away from rain Transport package shall not be exposed to sunlight Indicates correct upright position of the transport package Contents of the transport package are fragile ...

Page 3: ... monitoring during medical emergencies or operations If the cuff pressure exceeds 40 kPa 300 mmHg the unit will automatically deflate Should the cuff not deflate when pressures exceeds 40 kPa 300 mmHg detach the cuff from the arm wrist and press the button to stop inflation To avoid measurement errors carefully read this manual before using the product The equipment is not AP APG equipment and not...

Page 4: ... time The old battery is harmful to the environment so please do not dispose with other daily trash Remove the old battery from the device and follow your local recycling guidelines Setting User Date time and measurement unit It is important to set the clock before using your blood pressure monitor so that a time stamp can be assigned to each record that is stored in the memory 1 Switch User When ...

Page 5: ...meter 3 Set the Voice On Off After set the Date Time press SET button to enter the Voice mode Press MEM button to choose the Voice On or Off 4 5 After the Voice set press SET to enter the Unit mode Press MEM button to choose PA unit or mmHg unit Tie the cuff 1 Tie the cuff on your upper arm position the tube off center toward the inner side of arm in line with the little finger ...

Page 6: ...culation to recover For a meaningful comparison try to measure under similar conditions For example take daily measurements at approximately the same time on the same arm or as directed by a physician Start the Measurement 1 Press the On Off button to run the BP monitor it will begin to inflate and measure Blood Pressure automatically Do not move when in the measuring process The measurement will ...

Page 7: ... Press the MEM again it will show the latest 1 measurement and press again the second then the third Delete Memory 1 Press MEM button to choose the memory that needs to be deleted 2 Press SET button when the display shows dEL 3 Press MEM to confirm delete this memory 4 When the unit is off press MEM button for 3 seconds then press SET button until the device shows ALL EE that s means the memories ...

Page 8: ...r emotion exercise etc The attention need to pay When you measure you blood pressure at home If the cuff is tied properly If the cuff is too tight or too loose If the cuff is tied on the upper arm If you feel anxious pressured You had better take deep breath 2 3 times before beginning Advice adjust yourself for 4 5 minutes until you calm down ...

Page 9: ...ies Show on the display Batteries are low Replace with new batteries Error massage E 1 shows The heart beat signal is too weak or the pressure of cuff descend Refasten the cuff and then measure again E 2 shows The monitor detected motion while measuring Refasten the cuff and then measure again E 3 shows Measurement is not right Relax for a moment and then measure again E P shows Inflation fail Rel...

Page 10: ...AES U181 according to the EMC information provided in the accompanying documents Portable and mobile RF communications equipment can affect medical electrical equipment The use of accessories transducers and cables other than those specified with the exception of transducers and cables sold by MY WINGS as replacement parts for internal components may result in increased Emissions or decreased Immu...

Page 11: ...d are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class B The AES U181 is suitable for use in all establishments including domestic establishments Harmonic emissions IEC 61000 3 2 Not applicable Voltage fluctuations flicker emissions IEC 61000 3 3 Not applicable Table 202 Guidance and manufacturer s declaration electromagnetic immunity The AES U181 is ...

Page 12: ...n UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 sec Not applicable Mains power quality should be that of a typical commercial or hospital environment If the user of the AES U181 requires continued operation during power mains interruptions it is recommended that the AES U181 be powered from an uninterruptible power supply or a battery Pow...

Page 13: ...ixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and...

Page 14: ...ng to frequency of transmitter Electrical fast transient burst IEC 61000 4 4 150 kHz to 80 MHz d 1 2 p 80 MHz to 800 MHz d 1 2 p 800 MHz to 2 5 GHz d 2 3 p 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable ...

Reviews: