Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines ±1 kV for
input/output
lines
Not applicable
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±1 kV differential
mode
±2 kV common mode
Not applicable
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5 % UT (>95 % dip
in UT) for 0,5 cycle
40 % UT (60 % dip in
UT) for 5 cycles
70 % UT (30 % dip in
UT) for 25 cycles
<5 % UT (>95 % dip
in UT) for 5 sec
Not applicable
Mains power quality should be
that of a typical commercial or
hospital environment. If the
user of the AES-U181 requires
continued operation during
power mains interruptions, it is
recommended that the
AES-U181 be powered from
an uninterruptible power
supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.