Precaution For Use

 

Thanks you for purchasing our 

MEWOI1310B Large Caliber Wireless H/L  Voltage U-Type 

Current Meter

. In order to better for use of the product, please be certain:     

——Read  this  user  manual  in  detail,  the  operators  must  fully  understand  the  manual 

instructions and be able to operate the instrument skillfully, and then can field testing. 

——Strictly comply with the safety rules and precautions listed in this manual. 

 

 

Under any circumstance, shall pay special attention on safety in using this meter. Especially 

when measuring voltage circuit over AC100V and above. 

 

If the tested line voltage over 600V, it must be used insulation rod to connection. 

 

Because the high voltage circuit is very dangerous, the operator must be strictly trained and 

obtain the national high voltage operation certification before using the meter for field test. 

 

It is strictly forbidden to use this meter to test bare wires or busbars with voltage over 500kV. 

 

Pay attention to the text labeled on the panel and backplane of the Meter. 

 

Do  not  place  and  store  the  meter  in  high  temperature  and  humidity  or  dewy  places  and 

under direct sunlight for a long time.   

 

Replace the battery, please pay attention to the battery polarity. If you not use the leaker for 

a long time, please take out batteries. 

 

Disassembly,  calibration,  and  repair  of  this  tester  must  be  performed  by  authorized 

personnel.   

 

If meter`s components are damaged, Please stop to use.   

 

Due to the reason of this instrument, if it is dangerous to continue using, should stopped and 

sealed immediately ,and handled by an authorized institution. 

 

The  meter  manual  with  the  danger  mark  "        ",  users  must  follow  instructions  to  operate 

safely.   

 

The  meter  manual  with  the  extremely  dangerous  mark  "        ",  users  must  in  strict  follow 

instructions to operate safely.     

 

It is recommended that the meter should make insulation strength test at least once a year 

(AC 220kV/ RMS telescopic insulation rod is fully opened between the two ends). 

 
 
 
 
 
 
 
 
 
 

Summary of Contents for MEWOI1310B

Page 1: ...Wireless High Voltage U Type Current Meter MEWOI1310B User Manual www mewoi com...

Page 2: ......

Page 3: ......

Page 4: ...or Special Symbols 5 5 3 Display Demonstration 6 6 Operation Method 7 6 1 Detector Operation 7 6 1 1 Power On Off 7 6 1 2 General Test 7 6 1 3 Test Area 8 6 1 4 Data Transmission 8 6 2 Receiver Operat...

Page 5: ...o use this meter to test bare wires or busbars with voltage over 500kV Pay attention to the text labeled on the panel and backplane of the Meter Do not place and store the meter in high temperature an...

Page 6: ...se Below 600V circuit no need the insulation rod and use the u type current meter to measure directly The meter has data hold and data storage function Insulation rod is equipped with adjustable inter...

Page 7: ...tage AC current on line monitoring of low voltage AC current and AC current Power Supply DC6V Alkaline Dry Battery 1 5V AAA X4pcs Test Method Non contact U type CT Transmission Method 433MHz wireless...

Page 8: ...ement range overflow function OL A symbol display No signal Indication When the receiver has not received transmit signal dynamic display no symbol Auto Shut Down About 15 minutes after boo up the met...

Page 9: ...symbol 5 1 4 Date access symbol 5 1 5 2 digit Storage Data Serial Number 5 1 6 Unit Symbol 5 1 7 Date Locked Symbol 5 1 8 Decimal System Radix Point 5 1 9 4 Digit LCD Digital Display 5 2 Explanation...

Page 10: ...meter not in test mode or adjust received signal location and distance 5 3 Display Demonstration 5 3 1 The tested current 0 002A 2mA 5 3 2 Locked display Data The data is automatic stored as group 3...

Page 11: ...ric shock causing personal injury or injury accident Dangerous It is strictly forbidden to use this meter to test bare wires or busbars with voltage over 110kV otherwise there would be danger of elect...

Page 12: ...ill increase by twice or more and the position error of the semicircle area at point B will increase by 1 6 1 4 Data Transmission The meter has the function of wireless transmitting test data When the...

Page 13: ...l indication Press HOLD key again to cancel data lock state return to receiving data mode HOLD symbol disappear 6 2 4 Data Storage In the mode of receiving data press the HOLD key to hold the data whi...

Page 14: ...ace batteries in time 7 2 Power off and make sure the meter is off state Loosen two screws and open the battery cover plate to replace new qualified batteries pay special attention on the battery spec...

Page 15: ...11...

Page 16: ...12 Your Testing Specialist...

Reviews: