
44
9.4 480B 50/60Hz Power Line Locating
When locating a power line the 480B Transmitter is not required as you are locating the
power line by tracking the path of the 50/60 current coming off the line (the line must be
loaded).
This method gives a maximum (loud tone) signal. The 480B Receiver is held in a vertical
position at right angles to the line for maximum signal. (Note the logical path of the line from
the source).
You will notice a difference audio tone when using the 50/60Hz mode, this is normal.
50/60 Power Line Locating Operation:
1 Move the 480B Receiver a few feet away from where you want to start searching.
2 Turn the Receiver MODE Switch to “Passive 50/60Hz”.
3 Hold the Receiver in a vertical position. Then adjust the SENSITIVITY knob to
bring the needle within the “SET” zone. If the meter does not adjust to “SET”, turn
the SENSITIVITY knob all the way to the right (clockwise) and/or move closer to the
conductor. If you still cannot adjust to the “SET” position, switch the RANGE knob
to “HIGH”:, and then attempt to adjust the SENSITIVITY to “SEY”. (If no current is
present you will be unable to adjust to “SET”).
4 Hold the Receiver in an upright, vertical position with the face of the Receiver facing
you, and at right angles to the targeted conductor.
5 Move the Receiver from side to side over the targeted conductor location until you
obtain the maximum PEAK response. For better signal reception, hold the Receiver
close to the ground. As you move closer to the targeted conductor, the speaker tone
will gradually get louder. The loudest tone will indicate you are over the conductor in
the PEAK mode.
6 If the needle is at or close to the end of the meter scale, adjust the SENSITIVITY
knob to bring the needle back to the middle of the meter. (Decreasing the
SENSITIVITY will give a sharper Receiver response as you move closer to the target
conductor.) Then move to a new position and note the meter response. The meter
reading will increase if the signal gets stronger (closer to the conductor), or decrease
is the signal gets weaker (further from the conductor).
7
Keep adjusting the SENSITIVITY knob and the position of the Receiver until you
find a place that gives a maximum reading. (This will occur when the Receiver forms
an angle of 90 degrees with the target conductor.)