Figure 9-1: Direct (Conductive) Connection
4 Set the MODE switch to Line ONLY Locate. The Locate Power LED should be flashing
steadily, indicating the output signal is coming out.
5 Set the Receiver MODE switch to “Active”.
6 Set the Receiver SENSITIVITY KNOB to a low setting on the left of the dial, adjusting
as needing for good signal reception. As you move farther away from the Transmitter,
you will need to increase the sensitivity. It is important to tune the sensitivity as low as
possible in order to sharpen the reception and reduce signal bleed-off onto adjacent
conductors.
7 Set the Receiver RANGE knob to “NORM”, adjusting to “HIGH” when necessary.
8 When starting a locate, you need to “pre-locate” the targeted conductor by using the broad
range PEAK method first to find the general location of the line. Then hone in on the
exact location of the conductor with the precision NULL method.