![METROTEC GSM Series Operating Handbook Download Page 20](http://html.mh-extra.com/html/metrotec/gsm-series/gsm-series_operating-handbook_1780071020.webp)
METROTEC
20
Example of a display with trend in the range between 0 – 10
6.5 O
2
Red (O
2
in H
2
)
In order to obtain the oxygen value in a mixture which may for example consist of
nitrogen and hydrogen, the oxygen partial pressure can be used to calculate this
value based on the hydrogen proportion.
For the calculation the hydrogen proportion from menu line 26 is used.
Note:
The calculation of O
2
Red is a mathematical function. If the hydrogen proportion
changes or is not present, this value can not be calculated correctly.
Example of a display with trend in the 0 – 1000 ppm range