7.1 Remote interface
■■■■■■■■■■■■■■■■■■■■■■
32
■■■■■■■■
890 Titrando
Assignment
Pin No.
Function
*
Output 8
6
Output 9
7
Output 10
8
Output 11
13
Output 12
19
Output 13
20
0 volts / GND
14
+5 volts
15
0 volts / GND
25
* Signal activated only for operation with Touch Control.
Table 2
Explanation of the individual functions
Function
Explanation
Start
The current method is started at the time of acti-
vation.
t
pulse
> 100 ms
Stop
The current method is canceled (Stop) at the time
of activation.
t
pulse
> 100 ms
Quit
The current command in the determination run
will be canceled at the time of activation.
t
pulse
> 100 ms
Ready
The instrument is ready to receive a start signal.
Conditioning OK
The line is set when Conditioning with SET titra-
tion and KFT titration is at OK. The line remains
set until the determination is started with
[START]
.
Determination
The instrument performs a data-generating
determination.
EOD
E
nd
o
f
D
etermination.
Pulse (t
pulse
= 200 ms) after a determination or
after a buffer/standard solution during calibration
using a Sample Processor.
Error
The line is set for error message display.