Smart tips for Mira instruments
■■■■■■■■■■■■■■■■■■■■■■
20
■■■■■■■■
7 Operation and control
7.1
Smart tips for Mira instruments
The following smart tips are available.
Smart tips are attached to the instrument with magnetic connectors. The
smart tips contain a memory chip so that they can be identified by the
instrument. Due to its design, smart tips cannot be attached in a wrong
position.
Laser Classification of smart tips is described here
(see Chapter 2.4.3,
page 8)
.
The scope of delivery depends on your instrument version
(see Chapter
1.2, page 1)
.
1
Long distance attachment lens (LWD)
A long distance lens for point and shoot
measurement with laser class 3.
The focal point is approximately
8 mm
from
the top of the lens.
The LWD lens is used for samples in thick-
walled bottles.
2
Short distance attachment lens (SWD)
A short distance lens for point and shoot
measurement with laser class 3.
The focal point is approximately
0.85 mm
from the top of the lens.
The SWD lens is used for samples with direct
contact or in thin plastic bags.