5 Operation and maintenance
930 Compact IC Flex SeS (2.930.1400)
■■■■■■■■
83
5.10.3.5
Replacing parts of the Metrohm Suppressor
Module (MSM)
Parts of the Metrohm Suppressor Module (MSM) may need to be replaced
in the following cases:
■
Irremediable loss of suppressor capacity (reduced phosphate sensitivity
and/or significant rise in the baseline).
■
Irremediable blockage of the MSM (solutions can no longer be pum-
ped through the MSM).
Both the rotor and the connecting piece can be replaced.
Replacing parts of the Metrohm Suppressor Module (MSM)
1 Disconnecting the Metrohm Suppressor Module (MSM) from
the IC system
■
Switch off the instrument.
■
Disconnect all capillaries of the MSM from the IC system.
2 Dismantling the Metrohm Suppressor Module (MSM)
■
Unscrew the union nut
from the housing
■
out of the housing together with
the rotor
.
If the rotor gets stuck in the housing, you can push it out as fol-
lows:
Put a pointed object into the slot in the housing and use it to
push out the rotor.
■
Loosen the connecting piece from the rotor with a turning move-
ment.
3 Cleaning the new rotor
■
Clean the sealing surface of the new rotor
using a lint-free cloth.
4 Inserting the new rotor
CAUTION
The rotor may be destroyed during start-up if it is not inserted cor-
rectly.