![Metrohm 930 Compact IC Flex Manual Download Page 85](http://html.mh-extra.com/html/metrohm/930-compact-ic-flex/930-compact-ic-flex_manual_1779620085.webp)
■■■■■■■■■■■■■■■■■■■■■■
5 Operation and maintenance
930 Compact IC Flex SeS (2.930.1400)
■■■■■■■■
77
CAUTION
The Metrohm Suppressor Module (MSM) must be regenerated
(see Chapter 5.10.3.4, page 80)
or replaced
(see Chapter 5.10.3.5, page 83)
rohm Suppressor Module (MSM) is reduced or if the back pressure is
high.
5.10.2
Taking care of the suppressor housing
CAUTION
The transparent suppressor housing can become fogged.
The suppressor housing is made of PMMA (poly(methyl methacrylate)).
It can become scratched and foggy if cleaned improperly. This can make
viewing the rotor difficult or impossible.
■
Do
not use any scouring agents
for cleaning.
■
Do
not use any solvents
for cleaning.
5.10.3
Servicing the Metrohm Suppressor Module (MSM)
5.10.3.1
Parts of the Metrohm Suppressor Module
(MSM)
1
2
3
4
5
Figure 22
Parts of the Metrohm Suppressor Module (MSM)
1
Union nut
2
Connecting piece
3
Rotor
4
Housing
5
Slot in the housing