3 Installation
867 pH Module
■■■■■■■■
25
3.5.4
Connecting a temperature sensor or an electrode with integrated
temperature sensor
A temperature sensor of the Pt1000 or NTC type can be connected to the
Temp.
connector.
Connect the temperature sensor or the electrode with integrated tempera-
ture sensor as follows:
1
Insert the plugs of the temperature sensor into the
Temp.
sockets of
the pH Module.
Figure 15
Connecting a temperature sensor or an electrode with
integrated temperature sensor
NOTICE
Always insert the red plug into the red socket. This is the only way
that shielding against electrical interference can be ensured.
3.5.5
Connecting the iConnect
The external measuring interface 854 iConnect can be connected to the
iConnect
connector.
Connecting the 6.2168.000 mini USB adapter cable to an
instrument
If the mini USB adapter cable is not connected to the instrument yet, pro-
ceed as follows:
1
Plug the mini USB adapter cable (2) into the
iConnect
connector of
the instrument (1). Observe the correct orientation (markings).