6 System settings
848 Titrino plus
■■■■■■■■
57
Slope
This parameter is only visible with pH electrodes.
Slope of the pH electrode. With a 1-point calibration, only pH(0) can be
calculated, 100.0% is used as the slope.
Input range
–999.9 - 999.9 %
Default value
100.0 %
pH(0)
This parameter is only visible with pH electrodes.
pH value of the pH electrode at 0 mV. Apart from the slope, pH(0) is the
second characteristic of the calibration curve.
Input range
–20.000 - 20.000
Default value
7.000
Calibration temp.
This parameter is only visible with pH electrodes.
Temperature at which the last calibration was carried out.
Input range
–20.0 - 150.0 °C
Default value
25.0 °C
Calibration date
This parameter is only visible with pH electrodes.
Date of the last calibration.
Monitoring
This parameter is only visible with pH electrodes.
Activating and deactivating the calibration monitoring.
Selection
on
|
off
Default value
off
Time interval
This parameter is visible only when
Monitoring
=
on
.
You will be notified that this time interval (in days) has elapsed when start-
ing a method. You can then select whether or not you would still like to
start the method.
Input range
1 - 999 d
Default value
999 d