4.1
Technical data
Connector
Dsub connector 9 pin (male)
1
5
9
6
Baud rate
1'200
…
115'200 (für Dev. 8 1'200
…
19'200)
Data bits
7, 8
Stop bits
1, 2
Handshake
Xon/Xoff, none
Parity
none, even, odd
FIFO
0
…
16 Bytes (für Dev. 8 not available)
Pin assignment
Pin 2:
RxD (Received Data) for Dev. 1
…
4
Pin 3:
TxD (Transmitted Data) for Dev. 1
…
4
Pin 5:
GND (Signal Ground) for Dev. 1
…
4
Pin 7:
TxD (Transmitted Data) for Dev. 5
…
8
Pin 8:
RxD (Received Data) for Dev. 5
…
8
Pin 9:
GND (Signal Ground) for Dev. 5
…
8
Output lines
7 potential-free relay contacts per system for
controlling external instruments.
RUN
COM
Maximum contact load:
30 V / 1 A DC
Input lines
1 digital input per system for external method
start.
+5V
470k
47k
Ext. Start
RUN
COM
Maximum load for external contact:
approx. 0.1 mA at max. cut-off voltage of approx.
5 V